
AVTMDELTA4000-EN Rev 6 April 2018
63
9
Maintenance and Calibration
Maintenance
Maintenance should be performed only by qualified persons familiar with the
hazards involved with high-voltage test equipment. Read and understand Section
2,
Safety
, before performing any service.
Routine maintenance is all that is required for these test sets. The cables and
connector panel should be inspected frequently to be sure all connections are
tight and all ground connections intact.
DISCONNECT the MAINS plug before cleaning.
The appearance of the test set can be maintained by occasional cleaning of the
case, panel and cable assemblies. The outside of the carrying case can be cleaned
with detergent and water. Dry with a clean, dry cloth. The control panel can be
cleaned with a cloth dampened with detergent and water. Do not allow water to
penetrate panel holes, because damage to components on the underside may
result. A household all-purpose spray cleaner sprayed on cloth can be used to
clean the panel. Polish with a soft, dry cloth, taking care not to scratch the
display screen cover. The cables and mating panel receptacles can be cleaned
with isopropyl or denatured alcohol applied with a clean cloth.
Contamination of some parts of the high-voltage circuit, in particular the high-
voltage cable terminations and its mating panel receptacle, may show up as a
residual PF(DF) meter reading. Cleaning of these sensitive parts will remove the
leakage paths which cause the unwanted leakage current. Treat the high-voltage
cable with care. Keep it clean and do not subject it to abuse, such as dropping or
crimping.
Calibration
During the warranty period, no calibration should be necessary. Contact the
factory if there is any suspected problem.
The overall accuracy of capacitance and power factor (dissipation factor) at 10
kV should also be checked at least once a year against Megger’s Capacitance and
Dissipation Factor Standard (Cat. No. 670500-1). This will ensure that the entire
high-voltage circuit is functioning and calibrated properly.
Summary of Contents for DELTA4000
Page 2: ......
Page 3: ...DELTA4000 12 kV Insulation Diagnostic System ...
Page 8: ......
Page 14: ...M AVTMDELTA4000 EN Rev 6 April 2018 6 M ...
Page 28: ...M AVTMDELTA4000 EN Rev 6 April 2018 20 M ...
Page 58: ...M AVTMDELTA4000 EN Rev 6 April 2018 50 M ...