DC Insulation Resistance and HiPot Test Specifications
Table 18:
DC IR and HiPot test specifications.
Parameter
Variant
Value
Voltage accuracy
± 2% ± 5 V
Maximum output current
1.2 mA
Displayed current resolution
1 nA
Current measurement resolution
16 pA
Current accuracy
Test voltage 0–2 kV
± 4% ± 5 nA
Test voltage 2–4 kV
± 4% ± 10 nA
Test voltage 4–8 kV
± 4% ± 25 nA
Over current trip settings
Adjustable to 1.2 mA
IR measurement range
100 kΩ–1 TΩ
Surge Test Specifications
Table 19:
Surge test specifications.
Parameter
Variant
Value
Nominal surge capacitance
100 nF
Typical surge energy
11.25 J at 15 kV
Typical short circuit current
700 A
Repetition rate
4 Hz nominal
Minimum inductance
4 kV
70 µH
6 kV
100 µH
12 kV
120 µH
15 kV
170 µH
Voltage accuracy
±10%
Table 20:
Surge through Impedance Transformer test specifications.
Parameter
Value
Maximum voltage
1.5 kV
Short circuit current
2000 A
Minimum inductance
3 µH
Voltage accuracy (Voltages < 1 kV)
±12% ±3 V
Table 21:
Surge with Partial Discharge (PD) test specifications.
Parameter
Value
Inception and extinction voltages (PDIV, PDEV)
Measured per IEC 61934
Repetitive inception and extinction voltages (RPDIV, RPDEV)
Measured per IEC 61934
Programmable PD threshold range (Resolution to 0.1 mV)
1.0–999 mV
PD time scaling
1.024–26,400 µs
www.megger.com
Megger Baker ADX User Guide
167
DC Insulation Resistance and HiPot Test Specifications
Summary of Contents for ADX15
Page 1: ...Megger Baker ADX Automated Diagnostic Test Equipment User Guide...
Page 2: ...www megger com Megger Baker ADX User Guide...
Page 10: ...www megger com Megger Baker ADX User Guide viii Table of Contents...
Page 44: ...www megger com Megger Baker ADX User Guide 34...
Page 84: ...www megger com Megger Baker ADX User Guide 74 Routes...
Page 148: ...www megger com Megger Baker ADX User Guide 138...
Page 172: ...www megger com Megger Baker ADX User Guide 162...