Error and Failure Messages List
Table 7:
Description of error messages.
Message
Description
Appears in
Test Lead: ___ has failed due to the
E-Stop being Activated, would you like
to retest?
User has engaged the E-stop on the
front panel, after disengaging it this
error message is displayed.
Manual Surge, Automatic Surge, Coil
Test
Test Lead: __ has failed due to PPEAR+
limit exceeding the set limit in test
configuration, would you like to
retest?
The inputed PPEAR limit from the test
config the user had selected for the
test has been exceeded resulting in a
failure.
Manual Surge, Automatic Surge, Coil
Test
Pulse-to-Pulse EAR+ limit has been
exceeded
The active test config's PPEAR limit has
been exceeded during testing resulting
in a failure. Stops on failure.
Manual Surge, Automatic Surge, Coil
Test
Line-to-Line EAR+ limit has been
exceeded
The active test config's LLEAR limit has
been exceeded during testing resulting
in a failure.
Manual Surge, Automatic Surge, Coil
Test
The current Voltage has exceeded the
Target Voltage by over 10%
During testing the voltage has gone
over the target by 10% resulting in
test failure.
Manual Surge, Automatic Surge, Coil
Test
No leads have been tested yet, no data
to save
User has hit save on the screen
without testing.
Manual Surge, Automatic Surge, Coil
Test
___ is missing Calibration, Please
Calibrate to be able to use.
Unit has not been calibrated for host
for Surge, without having at least host
calibrated unit will not allow for Surge
Testing.
Manual Surge, Automatic Surge, Coil
Test
Power Pack is in the wrong mode,
please set to Surge mode
If power pack is set to incorrect mode
and must be placed to Surge to work
with ADX. ().
Manual Surge, Automatic Surge, Coil
Test
Input Value must be a number, Go to
Coil #
User has inputed something other
than a number (ex: nine instead of 9)
into the search field for the coil they
want to be set at.
Coil Test
Attribute must have a Name
Optional Attribute field for Active Test
Config screen before Auto Testing.
All Single Automatic Tests
Attribute must have a Value
Optional Attribute field for Active Test
Config screen before Auto Testing.
All Single Automatic Tests
Name and Value cannot be the same
Optional Attribute field for Active Test
Config screen before Auto Testing.
All Single Automatic Tests
Download Failed
Network issues causing the software
update to not fully download,
or causes corruption of file wile
downloading.
Software Update Button
APK for update not found
APK has been removed or renamed
under software table in Megger Server.
Software Update Button
Estop Activated!
Displays Dialog box letting test user
know that E-stop is engaged.
Automatic DC
Calibration Export Failed
ADX failed to export Calibration data. Calibration Export Screen
Calibration import Failed
ADX failed to import Calibration data. Calibration Import Screen
Unable to Save
User Config failed to save.
User Config
Unable to Create
User Config was unable to create.
User Config
Unable to Delete
User Config was unable to delete.
User Config
www.megger.com
Megger Baker ADX User Guide
142
Error and Failure Messages List
Summary of Contents for ADX15
Page 1: ...Megger Baker ADX Automated Diagnostic Test Equipment User Guide...
Page 2: ...www megger com Megger Baker ADX User Guide...
Page 10: ...www megger com Megger Baker ADX User Guide viii Table of Contents...
Page 44: ...www megger com Megger Baker ADX User Guide 34...
Page 84: ...www megger com Megger Baker ADX User Guide 74 Routes...
Page 148: ...www megger com Megger Baker ADX User Guide 138...
Page 172: ...www megger com Megger Baker ADX User Guide 162...