Insulation &

Low Resistance Tester

User Guide

SAFETY WARNINGS

Safety Warnings and Precautions must be read and understood

before the instrument is used. They must be observed during use.

The circuit under test must be switched  off, de - energized and 
isolated before Insulation or Continuity tests are made.

Circuit connections must not be touched during a test.

The test button must not be pressed while connecting the test 
leads or while changing ranges. 

The ‘Test’ button must not be pressed when making a voltage test.

The Default Voltmeter, and Automatic discharge are additional 
safety features and should not be regarded as a substitute for 
normal safe working practice.

The 

210170

is protected for connection to Power distribution 

systems up to 300 V Line Ground, and 500 V Line Line for 
Installation Category 

III

*

.

It is recommended that fused test leads are used when measuring 
voltage on high energy systems.

After insulation tests, capacitive circuits must be allowed to 
discharge before disconnecting the test leads.

Test leads, prods and alligator clips must be in good order; clean, 
and with no broken or cracked insulation.

Replacement fuses must be of the correct size, type and rating.

Symbols used on the instrument

Caution: risk of electric shock

Caution: refer to accompanying notes

Equipment protected throughout by Double 
Insulation (Class 

II

)

NOTE

THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT

PERSONS

.