Tests: Inductive mode
www.megger.com
DLRO2 and DLRO2X
28
6.2 Auto stop
1. Turn rotary switch to select the inductive mode position.
NOTE : To AUTO SAVE results click the
SAVE ( ) button
Refer to 5. Data storage (DLRO2X only) on page 22).
6.2.1 First test
Inductive test screen appears, ‘Inductive Mode’ shows in
secondary field for a short time then disappears.
NOTE :
Soft key 1 toggles MANUAL / AUTO.
2. Select AUTO on soft key 1.
3. Connect C1-C2 and P1-P2 to the instrument and the unit
under test.
C and P indicators show red background with
for no
continuity or green background with
for good continuity.
Proceed when both show
. The continuity indicators are
active during the test and will update if continuity is lost.
4. To start the test press the TEST button.
During the test results will live update on screen. The test will
stop automatically when the instrument determines that the
reading has been stable for long enough and is unlikely to
change significantly; the user can override the instrument and
stop the test at any time by pressing the TEST button.
The Difference Meter is active.
When the test is stopped the displayed result is static.