35
Sensor glucose is above 22.2 mmol/L (400mg/dL). (HIGH) and Sensor
glucose is below 2.2 mmol/L (40mg/dL). (LOW).
⚫
After the warm-up phase, the values before the first calibration are
replaced by
“BG”.
⚫
When the Sensor calibration expires, the reading values will be
underlined.
⚫
Calibration will be marked with a red dot “ ”.
⚫
The green square blocks in the graph background represent the Low and
High limits of glucose.
⚫
In the landscape screen display, tap the Home Key to return to the Home
Screen.
⚫
When data is being recovered after sensor-reconnection, you cannot
enter landscape screen by long-pressing the graph.
⚫
When you are only using the insulin pump as stand-alone system, the
sensor graph displays the blood glucose recorded, and BG inputs will be
displayed as red dots
“ ”.
3.3.6 Alarm Status
Some Alarm status remains even after the alarm is cleared. The following alarms
will be displayed on the Home Screen
in this case:
PATCH BATT DEPLETED, PUMP OUT OF RANGE, EMPTY RESERVOIR, PATCH
EXPIRED, OCCLUSION DETECTED, PATCH ERROR, AUTO OFF, EXCEEDS MAX TDD,
EXCEEDS MAX 1HR, PUMP BASE ERROR, LOW SUSPEND and PRE LOW SUSPEND.
For example: