10.7.3 Inducing VF with 50 Hz Burst
You may use ventricular 50 Hz Burst to induce VF. To induce VF, the 50 Hz Burst induction
delivers a rapid burst of VOO pacing pulses to the ventricle. You may specify the amplitude
and pulse width of these pulses, but the pacing interval is fixed at 20 ms.
As long as you press and hold the [50 Hz BURST Press and Hold] button on the programmer
screen, the device continues delivering the induction (up to a maximum of 10 s).
The ventricular 50 Hz Burst interface includes features to help simplify DFT testing. For
more information, see Section 5.5, “Performing ventricular defibrillation threshold tests”,
10.7.3.1 Considerations for inducing VF with 50 Hz Burst
Warning: Monitor the patient carefully when using an EP study function. Have an external
defibrillator ready for use when inducing any tachyarrhythmia. An induced tachyarrhythmia
may degenerate to ventricular fibrillation.
ATP Before Charging and ATP During Charging – ATP Before Charging and ATP During
Charging are automatically disabled for 30 s after delivery of a 50 Hz Burst induction. This
prevents ATP therapies from interfering with defibrillation threshold testing.
10.7.3.2 How to deliver a ventricular 50 Hz Burst induction
Figure 221. Ventricular 50 Hz Burst induction screen
1. Select Tests > EP Study.
2. Select 50 Hz Burst from the list of inductions and therapies.
3. If the Select Chamber dialog box appears, select [RV]. Otherwise, make sure that the
Chamber parameter is set to RV.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
439