Figure 215. Episode plot with Progressive Episode Therapies programmed to On
1 Plot option
2 Detection and therapy annotations
3 Programmed detection intervals
9.9 Optimizing charge time with Automatic Capacitor
Formation
Capacitor formation helps maintain quick charge times and prompt high-voltage therapy
delivery. Between formations, the capacitors gradually lose efficiency, which results in
longer charge times. To maintain quick charging for high-voltage therapies, the high-voltage
capacitors should be formed periodically.
9.9.1 System solution: Automatic Capacitor Formation
The Automatic Capacitor Formation feature ensures that the high-voltage capacitors are
fully formed on a regular basis.
9.9.2 Operation of Automatic Capacitor Formation
Automatic Capacitor Formation forms the capacitors fully at regular intervals. The
capacitors are fully formed by charging the capacitors to full energy and allowing the charge
to dissipate for at least 10 min. The system records the date and time of each formation.
After each formation, the feature schedules the next automatic capacitor formation. As the
battery depletes over the service life of the device, charge times gradually increase overall
irrespective of capacitor formation.
Capacitors can be formed either automatically or manually. For information about manually
forming the capacitors, see Section 10.6, “Testing the device capacitors”, page 434.