Finally, if the device delivers a ventricular cardioversion therapy, it adjusts the energy value
for the next cardioversion or defibrillation therapy to be equal to or greater than the energy
value of the last delivered therapy. For example, if the device detects a VT, delivers a 35 J
cardioversion, and redetects the episode as FVT, the next cardioversion therapy will also
deliver 35 J, even if the programmed value is 20 J.
Note: Progressive Episode Therapies operation does not cause the device to skip ATP
During Charging. However, if ATP Before Charging is enabled, the device skips the ATP
sequence before charging, and only delivers an ATP sequence during charging.
9.8.3 Programming Progressive Episode Therapies
Select Params icon
⇒ VF | Therapies…
⇒ Shared Settings…
▷ Progressive Episode Therapies
9.8.4 Evaluation of Progressive Episode Therapies
9.8.4.1 Treated VT/VF episode text
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
⇒ Text
The episode text for a treated ventricular tachyarrhythmia episode lists all the therapies the
device delivered during the episode. If Progressive Episode Therapies is enabled, this fact
is noted before the list of therapies.