8.7.4 Programming High Rate Timeout
Select Params icon
⇒ Detection (V.)…
⇒ High Rate Timeout…
▷ VF Zone Only
▷ All Zones
8.7.5 Evaluation of High Rate Timeout
8.7.5.1 Data - Arrhythmia Episodes screen
If a High Rate Timeout period expires during an episode, the stored EGM for that episode
includes the “HT” Decision Channel annotation at the point when the High Rate Timeout
period expired.
Figure 174. Episode EGM
If the High Rate Timeout–All Zones period expires during an episode, the Initial Type field
in the episode text will include the text, “High Rate Timeout–All Zones”.
If the High Rate Timeout–VF Zone Only period expires during an episode, the Initial Type
field in the episode text will include the text, “High Rate Timeout–VF Zone Only”.