7.6.4 Programming Rate Adaptive AV
Note: The TherapyGuide feature suggests parameter values based on information entered
about the patient’s clinical conditions. Parameter values for this feature are included. For
more information, see Section 4.9, “Using TherapyGuide to select parameter values”,
page 76.
Select Params icon
⇒ Pacing…
⇒ Paced AV…
▷ Rate Adaptive AV <On>
▷ Start Rate
▷ Stop Rate
▷ Minimum Paced AV
▷ Minimum Sensed AV
7.7 Adjusting PVARP to changes in the patient’s heart rate
A fixed value for the Post Ventricular Atrial Refractory Period (PVARP) may not provide the
optimal PVARP setting to meet the changing needs of the patient. At low heart rates, PVARP
should be long enough to prevent pacemaker-mediated tachycardia (PMT). At elevated
heart rates, PVARP should be short enough to avoid 2:1 block and promote AV synchrony.
For related information, refer to Section 7.1, “Sensing intrinsic cardiac activity”, page 201,
and Section 7.2, “Providing pacing therapies”, page 212.
7.7.1 System solution: Auto PVARP
Auto PVARP adjusts PVARP in response to changes in the patient’s heart rate or pacing
rate.
7.7.2 Operation of Auto PVARP
Auto PVARP is available when the pacing mode is programmed to DDDR, DDD, DDIR, DDI,
or an MVP mode (AAIR<=>DDDR, or AAI<=>DDD). Auto PVARP functions when the device
is operating in the DDDR, DDD, DDIR, or DDI mode.
In a tracking mode (DDDR or DDD), Auto PVARP adjusts PVARP based on the current heart
rate of the patient. When the heart rate is low, PVARP is longer to prevent PMT. As the heart
rate increases, PVARP shortens to maintain 1:1 tracking. Auto PVARP allows 1:1 tracking
of atrial events up to 30 bpm above the heart rate or 100 bpm, whichever is greater.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
253