2.4 Lead evaluation and lead connection
Refer to the lead technical manuals for specific instructions and precautions about lead
handling.
Hex wrench – Use only the torque wrench supplied with the device. The torque wrench is
designed to prevent damage to the device from overtightening a setscrew. Other torque
wrenches, (for example a blue-handled or right-angled hex wrench) have torque capabilities
greater than the lead connector can tolerate.
Lead connection – Consider the following information when connecting the lead and the
device:
●
Cap abandoned leads to avoid transmitting electrical signals.
●
Plug any unused lead ports to protect the device.
●
Verify lead connections. Loose lead connections may result in inappropriate sensing
and failure to deliver arrhythmia therapy.
Lead Impedance – Consider the following information about lead impedance when
evaluating the lead system:
●
Ensure that the defibrillation lead impedance is greater than 20 Ω. An impedance of
less than 20 Ω may damage the device or prevent delivery of high-voltage therapy.
●
Before taking electrical or defibrillation efficacy measurements, move objects made
from conductive materials, such as guide wires, away from all electrodes. Metal objects,
such as guide wires, can short circuit a device and lead, causing electrical current to
bypass the heart and possibly damage the device and lead.
Patch leads – Do not fold, alter, or remove any portion of a patch lead. Doing so may
compromise electrode function or longevity.
2.5 Device operation
Accessories – Use this device only with accessories, parts subject to wear, and disposable
items that have been tested to technical standards and found safe by an approved testing
agency.
Atrial Capture Management – Atrial Capture Management does not adjust atrial outputs
to values greater than 5.0 V or 1.0 ms. If the patient needs atrial pacing output greater than
5.0 V or 1.0 ms, manually program the atrial amplitude and pulse width. If a lead dislodges
partially or completely, Atrial Capture Management may not prevent loss of capture.
Battery depletion – Carefully monitor battery longevity by checking battery voltage and
replacement indicators. Battery depletion eventually causes the device to stop functioning.
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
25