Select Params icon
⇒ Pacing…
⇒ RV Amplitude…
▷ Atrial Capture Management
▷ Atrial Amplitude Safety Margin
▷ Atrial Minimum Adapted Amplitude
▷ RV Capture Management
▷ RV Amplitude Safety Margin
▷ RV Minimum Adapted Amplitude
⇒ Additional Parameters…
▷ Acute Phase Remaining
▷ Acute Phase Completed (read only)
7.5.7 Evaluation of Capture Management
7.5.7.1 Quick Look II
Select Data icon
⇒ Quick Look II
Threshold trends – The Quick Look II screen shows trends of average capture thresholds.
The threshold data is collected by the automatic daily threshold tests performed by Capture
Management. Select the Threshold [>>] button to view the Lead Trends and Capture
Threshold diagnostic screens.
Quick Look II Observations – If there are significant observations about ACM or RVCM,
they are shown in the Quick Look II Observations window.
For detailed information about viewing and interpreting all of the information available from
the Quick Look II screen, see Section 6.2, “Viewing a summary of recently stored data”,
7.5.7.2 Capture Threshold trends
Select Data icon
⇒ Device/Lead Diagnostics
⇒ Capture Threshold Trends
Medtronic
PROTECTA™ XT DR D314DRG
Clinician Manual
249