Figure 91. Threshold and safety margin curves
Pulse Width (ms)
Amplitude (V)
0.0
0.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
Target amplitude
Pacing threshold curve
Threshold measurement
Safety margin curve
7.5.3 Operation of Atrial Capture Management
Atrial Capture Management (ACM) is available when the pacing mode is programmed to
DDDR, DDD, or an MVP mode (AAIR<=>DDDR or AAI<=>DDD), and functions when the
device is operating in one of these modes. If ACM is programmed to the Monitor or Adaptive
setting, the device conducts a pacing threshold search to determine the atrial pacing
threshold. If ACM is programmed to the Adaptive setting, the device uses the atrial pacing
threshold to define a target amplitude and adjusts the pacing amplitude toward the target
amplitude. The target amplitude is based on the programmed settings for the Atrial
Amplitude Safety Margin and the Atrial Minimum Adapted Amplitude parameters.
Note: In the event of partial or complete lead dislodgment, ACM may not prevent loss of
capture.
Medtronic
PROTECTA™ XT DR D314DRG
240
Clinician Manual