Figure 63. Ventricular sensing with a true bipolar or an integrated bipolar lead
1 Sensing with a true bipolar lead and RV Sense Polarity programmed to Bipolar
2 Sensing with a true bipolar lead and RV Sense Polarity programmed to Tip to Coil
3 Sensing with an integrated bipolar lead and RV Sense Polarity programmed to either Bipolar or
Tip to Coil
7.1.2 Operation of sensing thresholds
The device automatically adjusts sensing thresholds after certain paced and sensed events
to help reduce the oversensing of T-waves, cross-chamber events, and pacing pulses. The
threshold adjustment depends on the type of event that precedes the adjustment. During
an automatic adjustment, the sensing threshold automatically increases, but it gradually
decreases toward the programmed sensitivity value, which is the minimum amplitude that
can be sensed. The threshold decrease is intended to be rapid enough to allow subsequent
low-amplitude signals to be sensed. Threshold adjustment with nominal settings is shown