Notes:
●
The Model 2696 InCheck Patient Assistant cannot communicate with the implanted
device if the device is already in a wireless telemetry session.
●
Patient-Activated Symptom Log entries are not collected when tachyarrhythmia
episodes, including SVTs, are being detected by the device.
●
If the patient uses the activator during an episode, the device check marks the episode
log entry and records the following in the episode text: “Patient Symptom detected
during episode”. However, a separate patient-activated record is not created.
6.8.3.1 Viewing the episode interval plot
When you first select an episode from the episode log, the programmer displays a graph
that plots the V-V and A-A intervals versus time, and indicates the following information:
●
programmed detection intervals
●
point of detection or detection withheld
●
point of onset for AT/AF
●
points of therapy delivery
●
point of episode termination
Figure 47. Episode Plot
1 Use this button to switch the y-axis between Interval and Rate.
Medtronic
PROTECTA™ XT DR D314DRG
170
Clinician Manual