![Medtronic PROTECTA XT DR D314DRG Clinician Manual Download Page 166](http://html.mh-extra.com/html/medtronic/protecta-xt-dr-d314drg/protecta-xt-dr-d314drg_clinician-manual_1763148166.webp)
6.7.5.3 Viewing the OptiVol event log
Select Data icon
⇒ Alert Events
⇒ OptiVol Events
The OptiVol event log reports the last 7 OptiVol Events. For OptiVol 2.0 Fluid Status
Monitoring, an event log entry is recorded the first time that the patient’s OptiVol 2.0 Fluid
Index has reached or exceeded the OptiVol Threshold.
6.8 Viewing Arrhythmia Episodes data and setting data
collection preferences
The system provides a clinically-oriented arrhythmia episode log that enables you to quickly
view summary and detailed diagnostic data for arrhythmia episodes. Episode information
is available in several formats, including interval plot diagrams, EGMs, and text summaries.
Various filtering tools are available to give you precise control over the types of data
displayed.
6.8.1 How to view the Arrhythmia Episodes data
Select Data icon
⇒ Clinical Diagnostics
⇒ Arrhythmia Episodes
6.8.2 Viewing the episode log
The episode log is displayed in the upper portion of the Data - Arrhythmia Episodes screen.
It provides the following summary information for the episodes currently being stored in
device memory:
●
type of episode
●
the number of ATP sequences delivered (if any)
●
the number of shocks or the energy delivered (if any)
●
whether the last therapy delivered was successful
●
the date, time, and duration of the episode
●
the average atrial and ventricular beats per minute
Medtronic
PROTECTA™ XT DR D314DRG
166
Clinician Manual
Summary of Contents for PROTECTA XT DR D314DRG
Page 2: ......
Page 9: ...Index 488 Medtronic PROTECTA XT DR D314DRG Clinician Manual 9 ...
Page 506: ......
Page 507: ......