background image

Operating Instructions

BATTERY SUPPORT SYSTEM 2

Summary of Contents for Battery Support System 2

Page 1: ...Operating Instructions BATTERY SUPPORT SYSTEM 2 ...

Page 2: ......

Page 3: ...BATTERY SUPPORT SYSTEM 2 OPERATING INSTRUCTIONS ...

Page 4: ...f this manual to update this vital tracking information Responsibility for Information It is the responsibility of our customers to ensure that the appropriate person s within their organization have access to this information including general safety information provided in Section 1 Medtronic Emergency Response Systems 11811 Willows Road Northeast Redmond WA 98052 2003 USA Telephone 425 867 4000...

Page 5: ...s 2 2 Unpacking and Inspecting 2 3 Controls Indicators and Connectors 2 3 Inserting Batteries 2 5 3 Battery Maintenance Battery Maintenance Warning 3 2 Battery Maintenance Guidelines and Procedures 3 2 Receiving New Batteries 3 4 Storing Batteries 3 4 Discarding Recycling Batteries 3 5 4 Battery Support System 2 Maintenance The Battery Support System 2 Service Indicator 4 2 Cleaning 4 2 Fuse Repla...

Page 6: ...e 2 2 Controls Indicators and Connectors 2 3 Figure 2 3 Battery Fuel Gauge 2 4 LIST OF TABLES Table 2 1 Controls Indicators and Connectors 2 3 Table 4 1 Troubleshooting Tips 4 3 Table 4 2 Replacement Parts and Accessories 4 4 Table 4 3 Battery Support System 2 Specifications 4 4 Table 4 4 LIFEPAK and FASTPAK Battery Specifications 4 5 ...

Page 7: ...upport System 2 Model Number s 3010035 complies with 93 42 EEC Medical Device Directive class I conformity assessed per Annex 7 This product complies with Safety EN 61010 1 1993 amd 1995 Class I Pollution degree 2 Installation Category II EMC EN 60601 1 2 1993 Redmond October 25 2004 Michael D Willingham Vice President Regulatory Affairs This declaration applies to CE marked devices produced after...

Page 8: ......

Page 9: ...ms Inc 1 SAFETY INFORMATION This section provides important information to help you safely operate the Battery Support System 2 Familiarize yourself with all of these terms warnings and symbols Terms page 1 2 General Warnings and Cautions 1 2 Symbols and Abbreviations 1 3 Document Conventions 1 4 ...

Page 10: ...power during patient care Medtronic has no information regarding the performance or effectiveness of its LIFEPAK defibrillator monitors if they are used with other manufacturers batteries or battery chargers Using other manufacturers batteries or battery chargers may result in device failure and may void warranty Use only Medtronic batteries and the appropriate Battery Support System Possible elec...

Page 11: ...attery Support System MIN 801807 or the two well Battery Charger MINs 9 00284 9 00288 and 801530 to charge and maintain a FASTPAK 2 battery will eventually result in an inaccurate battery charge level indicator Use only the Battery Support System 2 MIN 3010035 to charge and maintain FASTPAK 2 batteries Possible equipment damage The supplied power cord is used to connect the battery support system ...

Page 12: ... Canadian Standards Association certification for Canada and the United States Marking of conformity according to the Medical Device Directive 93 42 EEC Hz Hertz frequency V Volts A Amperes W Watts T Time lag fuse conforming to IEC 127 LED Light Emitting Diode NiCd Nickel Cadmium chemistry battery SLA Sealed Lead Acid chemistry battery MIN Manufacturer s Item Number CAT Catalog number used for pla...

Page 13: ...sponse Systems Inc 2 BASIC ORIENTATION This section provides a basic orientation to the Battery Support System 2 Battery Support System 2 page 2 2 The LIFEPAK and FASTPAK Family of Batteries 2 2 Unpacking and Inspecting 2 3 Controls Indicators and Connectors 2 3 Inserting Batteries 2 5 ...

Page 14: ...e Charge FASTPAK batteries in either the Battery Support System MIN 801807 the Battery Support System 2 MIN 3010035 or the LIFEPAK 12 defibrillator when it is powered by the AC or DC Power Adapter FASTPAK batteries require periodic conditioning and shelf life testing to optimize performance Condition and shelf life test FASTPAK batteries in the Battery Support System or Battery Support System 2 Ch...

Page 15: ...d product literature from the shipping container Carefully inspect the system and the power cord for any signs of damage during shipping Verify the receipt of everything noted in the packing list Save the shipping container and the foam inserts in case the system needs to be returned CONTROLS INDICATORS AND CONNECTORS Figure 2 2 Figure 2 3 and Table 2 1 provide an overview of the controls indicato...

Page 16: ...acle Receptacle for ac power cord 11 Fuses Two fuses help protect the Battery Support System 2 from current surges 12 Power Cord Cable for operation from an ac outlet Figure 2 3 Battery Fuel Gauge 13 Fuel Gauge Button Lights the fuel gauge indicators 14 Fuel Gauge Indicators One light flashing charge battery Two alternately flashing lights battery requires conditioning No lights battery has 0 char...

Page 17: ...r signs of damage or distortion 3 Align the battery so that the battery clip is over the pins in the battery well 4 Insert the end of the battery opposite the battery clip into the battery well 5 Firmly press the other end into the battery well until it clicks into place To remove the battery press the battery clip in and lift the battery out of the battery well CAUTION Failure to charge batteries...

Page 18: ......

Page 19: ...nc 3 BATTERY MAINTENANCE This section describes how to use the Battery Support System 2 to maintain LIFEPAK and FASTPAK batteries Battery Maintenance Warning page 3 2 Battery Maintenance Guidelines and Procedures 3 2 Receiving New Batteries 3 4 Storing Batteries 3 4 Discarding Recycling Batteries 3 5 ...

Page 20: ...ncreased and irreversible cell damage could occur Note Fully charge SLA batteries between uses If SLA batteries are not 100 recharged between uses sulfation lead sulfate buildup on electrode surfaces inside the battery can occur Sulfation reduces battery capacity and may result in premature battery failure To charge a battery 1 Insert a battery into a Battery Support System 2 battery well CHARGE L...

Page 21: ...or FASTPAK 2 batteries to complete the condition cycle 2 Press CONDITION for other FASTPAK batteries 3 READY lights when a battery is fully recharged The Battery Support System 2 automatically recharges a battery that passes the conditioning process Note If DISCARD lights remove the battery from use and discard recycle Note If a power failure occurs during battery conditioning the Battery Support ...

Page 22: ...f life test and is fully recharged The battery may then be returned to use Note If a power failure occurs during shelf life test the Battery Support System 2 interrupts the shelf life test and reverts to CHARGE mode once power is restored shelf life test may not have been completed Repeat the shelf life test Note If DISCARD lights remove the battery from use and discard recycle RECEIVING NEW BATTE...

Page 23: ...IFEPAK NiCd LIFEPAK SLA and FASTPAK 2 batteries have internal parameters that establish limits for useful life If these parameters are exceeded the Battery Support System 2 will indicate DISCARD when battery is inserted into the battery well To promote awareness of battery recycling Medtronic LIFEPAK and FASTPAK batteries are marked with one of these symbols When a LIFEPAK or FASTPAK battery reach...

Page 24: ......

Page 25: ...TERY SUPPORT SYSTEM 2 MAINTENANCE This section describes the Battery Support System 2 maintenance The Battery Support System 2 Service Indicator page 4 2 Cleaning 4 2 Fuse Replacement 4 2 Troubleshooting 4 3 Replacement Parts and Accessories 4 4 Specifications 4 4 Warranty Policy 4 6 Recycling Information 4 6 ...

Page 26: ...ort System 2 Service Manual In the USA call 1 800 442 1142 for technical consultation manuals or parts Outside the USA contact your local Medtronic service office CLEANING Clean the Battery Support System 2 case battery contacts batteries and ac power cord with a damp sponge or towel The following agents may be used Mild soap and water Quaternary ammonium compounds Isopropyl alcohol Peracetic pero...

Page 27: ...e and contact a qualified service representative CHARGE CONDITION does not light when a FASTPAK battery is installed Battery malfunction Install a different battery in the same battery well If CHARGE CONDITION lights discard recycle the original battery Battery Support System 2 malfunction Remove from use and contact a qualified service representative READY does not light after hours of charging B...

Page 28: ...nches x 3 8 inches Weight 8 2 kg 18 pounds excluding batteries Number of battery wells 3 Charging conditioning and shelf life test capability in all battery wells Power requirements 100 120 220 240 10 50 60 Hz 3 Hz Fuses Two fuses in the power input module 5 x 20 mm T 250V Low or High break capacity per IEC 127 2 sheet 1 or 2 such as Bussman GDC Littelfuse 218 or Schurter FST 100 120 Volt Input 1 ...

Page 29: ... Range 0 to 35 C 32 to 95 F LIFEPAK SLA Battery Type Sealed lead acid Weight 1 4 kg 3 0 lbs Voltage 12Vdc Capacity 2 5 amp hours Charge Time fully depleted battery 6 hours typical 12 hours maximum Conditioning Time 28 hours typical 56 hours maximum Charging Temperature Range 5 to 35 C 41 to 95 F Operating Temperature Range 0 to 50 C 32 to 122 F Long Term 1 day Storage Temperature Range 0 to 35 C 3...

Page 30: ...of the defect Use of other manufacturers batteries and accessories with Medtronic defibrillator monitor products may void Safety Agency Certifications and warranty RECYCLING INFORMATION Recycle the system at the end of its useful life Recycling Assistance The system should be recycled according to national and local regulations Contact your local Medtronic representative for assistance Preparation...

Page 31: ...tem 2 maintenance 4 1 Cleaning 4 2 Fuse replacement 4 2 Replacement parts and accessories 4 4 Service indicator 2 3 4 2 Specifications 4 4 Troubleshooting 4 3 Warranty policy 4 6 Battery Support System 2 service indicator 2 3 4 2 C Charge indicator 2 4 3 2 3 3 4 3 Charge level indicator 2 2 Charging batteries 3 2 Cleaning 4 2 Communication pins 2 4 Condition indicator 2 3 3 3 4 3 Conditioning sche...

Page 32: ...ertifications 4 6 Safety information 1 1 Document conventions 1 4 General warnings and cautions 1 2 Symbols and abbreviations 1 3 Terms 1 2 Schedule battery conditioning 3 3 Self diagnostic tests 4 2 Self discharge rate 3 3 Service indicator 2 3 4 2 Shelf life indicator 2 3 Shelf life indicator 4 3 Shelf life testing 2 2 3 3 SLA batteries 3 3 Specifications 4 4 Storing batteries 3 4 Sulfation 3 2 ...

Page 33: ...ess actual physical location for example 123 Main Street Third Floor Suite A 3 City State and Zip Code 4 A contact name and telephone number 5 Device part number and serial number The address to which this particular equipment was shipped is the current tracking location If this equipment is located somewhere other than the shipping address or you have purchased this equipment from someone other t...

Page 34: ......

Page 35: ......

Page 36: ...5 867 4000 Toll Free USA only 800 442 1142 Fax 425 867 4121 Internet www medtronic ers com www medtronic com Medtronic Europe S A Medtronic Emergency Response Systems Rte du Molliau 31 Case postale 84 1131 Tolochenaz Switzerland Telephone 41 21 802 7000 Fax 41 21 802 7900 MIN 3010515 003 CAT 26500 000168 ...

Reviews: