SURE-SCREEN

®

  

Quick Reference Guide

 

 

 

 

The device detects one or more of the following 
drug classes at the following concentrations: 

Amphetamine (AMP) 

300 ng/ml 

Barbiturates (BAR) 

200 ng/ml 

Buprenorphine (BUP) 

10 ng/ml 

Benzodiazepines (BZO) 

200 ng/ml 

Cocaine (COC) 

100 ng/ml 

Methadone (MTD) 

200 ng/ml 

Methamphetamine (mAMP) 

1000 ng/ml 

Opiates (OPI) 

100 ng/ml 

Oxycodone (OXY) 

100 ng/ml 

Phencyclidine (PCP) 

25 ng/ml 

Propoxyphene (PPX) 

300 ng/ml 

Cannabinoids (THC) 

40 ng/ml 

 

Note:  If LFAS adulterant strip is present (to the right of drug test strips), 
refer to the Color Comparator Chart to interpret results

.

 

 

 

 

 

 

 

Obtain urine. Urine level 
must be between the Min / 
Max lines on cup. If urine 
was refrigerated, bring to 
room temp and mix well 
before testing. 

Screw lid onto cup --

TIGHTEN FIRMLY. 

Tear open pouch and 
remove device. Align device 
and cup as shown. 

Place the device securely on 
to the lid---the parts should 
interlock easily and move 
freely clockwise.  

Rotate the device clockwise 
¼ of a turn (i.e. from 9 
o’clock to 12 o’clock) until it 
snaps into place. 

Place cup on its side as 
shown above to start urine 
flow. 

 

 

 

 
 

For examples at right: 

C = Control 
T1 Tests = AMP, THC, BZO 
T2 Tests = COC, OPI 

 

VALID: 

A reddish-purple line appears 
i

n the “C” area for each test. 

 

 

 

 

 

Example Sufficient URINE Present 

 

--about ¼ inch 

evenly distributed-- 

 

is all that is needed for the device to 
migrate properly.  More urine is NOT 
necessary!  Allow test device to rest 
undisturbed for 5 minutes.

 

INVALID: 

No control line 

“C” present.  

Retest urine with a new 
device.   
 

NEGATIVE: 

Any reddish-purple line visible at 5 min. 
in the test area is a negative result 
(even a faint or broken line). 
 
Example above:  all 5 test results are 
negative. 

PRELIMINARY POSITIVE: 

The absence of a line in the test area 
is a preliminary positive result. 
 
Example above:  the AMP 
(Amphetamine) test is a preliminary 
positive. 

 

 

 

 
Seal Properly if 
Transporting to Laboratory 
for confirmation: 

 

Tighten the lid 

 

Use dual seal provided  
with cup 

o

 

Seal sample port  

o

 

Seal vent hole 

 

 

For complete product information

 refer to the 

PACKAGE

 

I

NSERT

 

available at 

http://www.medtox.com/ProductTraining.aspx

  

or contact Technical Support at 1-877-643-5703.   

SURE-SCREEN

®   

12-panel Forensic Use Only

  

Package  

INSERT Rev. 6/10, 

P/N 102109

.  

 

MEDTOX Diagnostics, Inc.  •  1238 Anthony Road 

Burlington, NC 27215 

Phone: 1-877-457-

0613  •  Fax: 651-286-6222 

E-mail: 

[email protected]

 

 

READING THE TEST RESULTS 
Read Time:  5 minutes 

C = Control Line 
T1 = Test Line for drug 

name on top portion 

T2 = Test Line for drug 

name below  
(if no print on lower 
color, then no T2 
drug tested) 

 

Bring pouched device to 
room temp (18-25°C / 64-
77°F) before opening. 

 

For Forensic Use Only 
 
 

Instructions 