Installation - System and Accessory
16 - 95
18.
Attach the Display Control Unit mast cover (G), and reconnect the contrast tray (D).
19.
Attach the Upper mast cover (C) while aligning the Injector Head cable into the molded guides
in the mast cover piece. Leave enough slack to allow full movement of the Injector Head on
the upper mast.
20.
Loop the Display Control Unit and Injector Head cables and secure with the Cable loop as
shown in the figure below.
Figure 16 - 7: Pedestal Cable Routing
21.
In the Lower mast, attach the cable guide with the Display Control Unit and Injector Head
cables behind it and in the channel on the lower mast. Ensure the flange hole for the upper
mast screw is facing up.
22.
Attach the Lower Mast cover (E).
23.
Connect the Power Unit power cord to the inlet plug on the Power Unit.
16.3 Power Unit Installation
For U.S installations, equipment shall only be connected to Hospital Grade or Hospital Only outlets. This
section shows all of the Power Unit connection points, describes how to relocate the connections from
the top plate to the back plate, and shows how to assemble the Power Unit floor mount bracket,.
Electric Shock Hazard - Serious patient and/ or worker injury or death may result.
•
Use only power cord approved for use on Mark 7 Arterion.
•
For U.S installations, equipment shall only be connected to Hospital Grade or Hospital Only
outlets.
Electro-Mechanical Hazard - Equipment damage may result.
•
Do not block Power Unit Vents.
•
Installation clearance should be a minimum of 3 to 5 inches (8 to 13 cm).
NOTICE
Summary of Contents for Mark 7 Arterion
Page 1: ...Operation Manual READ BEFORE USING...
Page 2: ......
Page 8: ...MEDRAD Mark 7 Arterion Injection System vi...
Page 10: ...MEDRAD Mark 7 Arterion Injection System 1 2...
Page 16: ...MEDRAD Mark 7 Arterion Injection System 2 8...
Page 23: ...Symbols and Icons 3 15 This Way Up Not made with natural rubber latex...
Page 24: ...MEDRAD Mark 7 Arterion Injection System 3 16...
Page 36: ...MEDRAD Mark 7 Arterion Injection System 5 28...
Page 48: ...MEDRAD Mark 7 Arterion Injection System 8 40...
Page 58: ...MEDRAD Mark 7 Arterion Injection System 9 50...
Page 68: ...MEDRAD Mark 7 Arterion Injection System 10 60 Figure 10 5 ISI Enabled...
Page 80: ...MEDRAD Mark 7 Arterion Injection System 11 72...
Page 90: ...MEDRAD Mark 7 Arterion Injection System 14 82...
Page 146: ...MEDRAD Mark 7 Arterion Injection System 17 138...
Page 152: ...MEDRAD Mark 7 Arterion Injection System 18 144...
Page 156: ...MEDRAD Mark 7 Arterion Injection System A 148...
Page 161: ......