Q-Sense Technical Reference Manual
Proprietary
124 of 188
Parameter
Description
Ramp timeout (% of
duration)
If Ramp temperature continues to be detected beyond this
setting, an error message is displayed. The test is not
aborted.
Thermode enabled
If the use of probe type is not possible (for example,
because it is broken) its usage can be disabled by
deselecting the checkbox.
10.2.2
Safety
The safety tab presents the software temperature time limitations which are allowed according to
the program safety regulations. In addition, is presents a list of the self-tested parts, and built-in
test (BIT) parameters.
The values in this tab can only be read, and cannot be edited.
Figure 83: Hardware settings – Safety tab
10.2.3
Thermode
The thermode tab presents all the thermode data and calibration tables. Refer to the Thermode
Management section in the Service Manual for further explanation.
Additional available functions are:
Thermode calibration.
Thermode management.