Medline EVENCARE G3 Operator'S Manual Download Page 26

 26

EvenCare

®

G3

 

Blood Glucose System

Performing a Control Solution Test

We reserve the right to correct any errors. 
©2016 Medline Industries, Inc., One Medline Place, Mundelein, IL 60060 | 1-800-MEDLINE  | www.medline.com
EvenCare and Medline are registered trademarks of Medline Industries, Inc.

MKT1661989 / e16015 / PDF / 3

step

1

step

4

step

2

step

5

step

3

step

6

Insert test strip to  
turn on meter.

Compare result to the low 
or high range printed on 
test strip box.

Invert and gently shake 
the control solution bottle 
three times. Apply a drop 
onto a clean, dry, non-
absorbent surface such 
as a table top.

Eject and discard the 
used test strip. The meter 
will automatically turn off. 
Wipe the droplet clean.

Press     or     once to 
enter control solution 
mode for low or high. 
The “ctl” icon will display  
in the lower left corner of  
the LCD, indicating the 
meter is in control mode.

STEP 1

STEP 2

STEP 5

STEP 3

STEP 4

STEP 6

Blood Glucose Test Strips

For use with the EvenCare G3

Professional Blood Glucose

Monitoring System

50

Individually 

Foil Wrapped

Test Strips

Important

• Do not freeze.

• Storage temperature 39˚F – 86˚F.

• Keep away from sunlight.

• For in vitro diagnostic use only.

• Single use only.

• CAUTION: Consult accompanying 

   documents.

www.medline.com  

©2015 Medline Industries, Inc. 

Medline and EvenCare are registered 

trademarks of Medline Industries, Inc. 

Manufactured in Taiwan for Medline 

Industries, Inc., Mundelein, IL 60060 USA

1-800-MEDLINE RI15APB 

U.S. Patent # 7,655,128.  

Product is not made with 

Natural Rubber Latex.

Control Solution Range (mg/dL)

Reagent Composition

Each test strip contains per cm

2

 

of reaction zone approximately:
Glucose Oxidase (

A. niger

2.3%

Electron Shuttle 

25%

Non-Reactive Ingredients 

72.7%

MPH3550FL

Exp:

Level 1:

Level 3:

44-74

190-258

STEP 1

STEP 2

STEP 5

STEP 3

STEP 4

STEP 6

Blood Glucose Test Strips

For use with the EvenCare G3

Professional Blood Glucose

Monitoring System

50

Individually 

Foil Wrapped

Test Strips

Important

• Do not freeze.

• Storage temperature 39˚F – 86˚F.

• Keep away from sunlight.

• For in vitro diagnostic use only.

• Single use only.

• CAUTION: Consult accompanying 

   documents.

www.medline.com  

©2015 Medline Industries, Inc. 

Medline and EvenCare are registered 

trademarks of Medline Industries, Inc. 

Manufactured in Taiwan for Medline 

Industries, Inc., Mundelein, IL 60060 USA

1-800-MEDLINE RI15APB 

U.S. Patent # 7,655,128.  

Product is not made with 

Natural Rubber Latex.

Control Solution Range (mg/dL)

Reagent Composition

Each test strip contains per cm

2

 

of reaction zone approximately:
Glucose Oxidase (

A. niger

2.3%

Electron Shuttle 

25%

Non-Reactive Ingredients 

72.7%

MPH3550FL

Exp:

Level 1:

Level 3:

44-74

190-258

STEP 1

STEP 2

STEP 5

STEP 3

STEP 4

STEP 6

Blood Glucose Test Strips

For use with the EvenCare G3

Professional Blood Glucose

Monitoring System

50

Individually 

Foil Wrapped

Test Strips

Important

• Do not freeze.

• Storage temperature 39˚F – 86˚F.

• Keep away from sunlight.

• For in vitro diagnostic use only.

• Single use only.

• CAUTION: Consult accompanying 

   documents.

www.medline.com  

©2015 Medline Industries, Inc. 

Medline and EvenCare are registered 

trademarks of Medline Industries, Inc. 

Manufactured in Taiwan for Medline 

Industries, Inc., Mundelein, IL 60060 USA

1-800-MEDLINE RI15APB 

U.S. Patent # 7,655,128.  

Product is not made with 

Natural Rubber Latex.

Control Solution Range (mg/dL)

Reagent Composition

Each test strip contains per cm

2

 

of reaction zone approximately:
Glucose Oxidase (

A. niger

2.3%

Electron Shuttle 

25%

Non-Reactive Ingredients 

72.7%

MPH3550FL

Exp:

Level 1:

Level 3:

44-74

190-258

STEP 1

STEP 2

STEP 5

STEP 3

STEP 4

STEP 6

Blood Glucose Test Strips

For use with the EvenCare G3

Professional Blood Glucose

Monitoring System

50

Individually 

Foil Wrapped

Test Strips

Important

• Do not freeze.

• Storage temperature 39˚F – 86˚F.

• Keep away from sunlight.

• For in vitro diagnostic use only.

• Single use only.

• CAUTION: Consult accompanying 

   documents.

www.medline.com  

©2015 Medline Industries, Inc. 

Medline and EvenCare are registered 

trademarks of Medline Industries, Inc. 

Manufactured in Taiwan for Medline 

Industries, Inc., Mundelein, IL 60060 USA

1-800-MEDLINE RI15APB 

U.S. Patent # 7,655,128.  

Product is not made with 

Natural Rubber Latex.

Control Solution Range (mg/dL)

Reagent Composition

Each test strip contains per cm

2

 

of reaction zone approximately:
Glucose Oxidase (

A. niger

2.3%

Electron Shuttle 

25%

Non-Reactive Ingredients 

72.7%

MPH3550FL

Exp:

Level 1:

Level 3:

44-74

190-258

STEP 1

STEP 2

STEP 5

STEP 3

STEP 4

STEP 6

Blood Glucose Test Strips

For use with the EvenCare G3

Professional Blood Glucose

Monitoring System

50

Individually 

Foil Wrapped

Test Strips

Important
• Do not freeze.

• Storage temperature 39˚F – 86˚F.

• Keep away from sunlight.

• For in vitro diagnostic use only.

• Single use only.

• CAUTION: Consult accompanying 

   documents.

www.medline.com  

©2015 Medline Industries, Inc. 

Medline and EvenCare are registered 

trademarks of Medline Industries, Inc. 

Manufactured in Taiwan for Medline 

Industries, Inc., Mundelein, IL 60060 USA

1-800-MEDLINE RI15APB 

U.S. Patent # 7,655,128.  

Product is not made with 

Natural Rubber Latex.

Control Solution Range (mg/dL)

Reagent Composition

Each test strip contains per cm

2

 

of reaction zone approximately:
Glucose Oxidase (

A. niger

2.3%

Electron Shuttle 

25%

Non-Reactive Ingredients 

72.7%

MPH3550FL

Exp:

Level 1:

Level 3:

44-74

190-258

STEP 1

STEP 2

STEP 5

STEP 3

STEP 4

STEP 6

Blood Glucose Test Strips

For use with the EvenCare G3

Professional Blood Glucose

Monitoring System

50

Individually 

Foil Wrapped

Test Strips

Important
• Do not freeze.

• Storage temperature 39˚F – 86˚F.

• Keep away from sunlight.

• For in vitro diagnostic use only.

• Single use only.

• CAUTION: Consult accompanying 

   documents.

www.medline.com  

©2015 Medline Industries, Inc. 

Medline and EvenCare are registered 

trademarks of Medline Industries, Inc. 

Manufactured in Taiwan for Medline 

Industries, Inc., Mundelein, IL 60060 USA

1-800-MEDLINE RI15APB 

U.S. Patent # 7,655,128.  

Product is not made with 

Natural Rubber Latex.

Control Solution Range (mg/dL)

Reagent Composition

Each test strip contains per cm

2

 

of reaction zone approximately:
Glucose Oxidase (

A. niger

2.3%

Electron Shuttle 

25%

Non-Reactive Ingredients 

72.7%

MPH3550FL

Exp:

Level 1:

Level 3:

44-74

190-258

CAUTION:

 Failure to follow instruction may cause out of range results.

NOTE:

• Write down the date when the control solution is first opened. Discard three months after the date of the first opening.
•  For more information on performing a control solution test, see the EvenCare G3 User’s Guide or call Customer Support 

at 

1-866-274-2131

.

Bring meter and strip to the 
droplet. The meter will beep 
and display count down bars 
after enough control solution 
is drawn into the test strip.

Summary of Contents for EVENCARE G3

Page 1: ...BLOOD GLUCOSE MONITORING SYSTEM Healthcare Professional Operator s Manual Operator s Manual In Service Guide...

Page 2: ...BLOOD GLUCOSE MONITORING SYSTEM Healthcare Professional Operator s Manual Operator s Manual In Service Guide...

Page 3: ...RTIFICATION CHECKLIST 17 SELF TEST 18 SELF TEST ANSWER KEY 19 EVENCARE G3 BLOOD GLUCOSE MONITORING SYSTEM NURSE IN SERVICE TRAINING RECORD 20 BLOOD GLUCOSE MONITORING SYSTEM CERTIFIED INSTRUCTOR RECOR...

Page 4: ...he quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips forearm or palm It is indicated to be used for multiple patients in a clinical setting by health...

Page 5: ...ne 3V Lithium coin cell battery CR2032 Test Strip Ejector Slide to eject used test strip Enter Button Press and hold to turn the meter on or off Data Port Function is not yet available Test Strip Port...

Page 6: ...trol Indicates a control solution test Test Strip Indicates the meter is waiting for the test strip to be inserted Blood Drop Appears when a test strip is inserted into the meter and the meter is read...

Page 7: ...e Level 3 Glucose Control Solution blue cap helps to validate system performance at a high range blood glucose reading Run the control test to make sure the meter and test strips are working properly...

Page 8: ...screen Step 6 For vial test strips compare the reading on the screen to the control solution range printed on the test strip bottle For foil wrapped test strips compare the reading on the screen to t...

Page 9: ...p and lance the finger Follow the manufacturer s instructions for how the lancing device should be used Step 4 When the blood drop icon appears on the screen gently bring the test strip to the test si...

Page 10: ...s shown on the screen Press the down button to see the next result and the next and so on Press and hold the up button will allow you to flash through the results in the meter nil will appear on the d...

Page 11: ...2 Put on single use medical protective gloves Step 3 Inspect for blood debris dust or lint anywhere on the meter Blood and bodily fluids must be thoroughly cleaned from the surface of the meter Step 4...

Page 12: ...ing devices used in a clinical setting should be the safety models marketed for single use only Glucose meters used in a clinical setting for testing multiple persons must be cleaned and disinfected b...

Page 13: ...he test with a new test strip The strip was removed from the meter while testing was in progress Repeat the test with a new test strip The size of the blood sample was too small Repeat the test with a...

Page 14: ...nted to be free of defects in workmanship and materials for a period of two 2 years from the date of purchase Medline s liability for warranty claims is limited to repair or replacement and in no even...

Page 15: ...l resource should you choose to use it IN SERVICE GUIDE The success of any bedside glucose monitoring program depends on having consistent well documented methods and procedures for testing and record...

Page 16: ...er s Guide Toll free customer service number 2 Be familiar with the EVENCARE G3 test strip packaging and location of the lot number expiration date and control solution ranges 3 Be familiar with the p...

Page 17: ...and Disinfecting Storage 2 EVENCARE G3 Blood Glucose Test Strips Expiration Date Lot Number Control Solution Ranges Handling of Strip Proper Insertion of Strip Storage 3 EVENCARE G3 Control Solution E...

Page 18: ...ting it into the meter 4 True or False You need to mark a control test as a control test so that it shows as a control test in the memory of the meter 5 True or False For vial test strips the vial cap...

Page 19: ...test in the memory of the meter 5 True or False For vial test strips the vial cap should be closed immediately after removing a test strip from the vial to avoid any degrading or contamination of the...

Page 20: ...SYSTEM NURSE IN SERVICE TRAINING RECORD The healthcare professionals listed below have been trained to use the EVENCARE G3 Blood Glucose Monitoring System Certified Trainer s Name DATE HEALTHCARE PROF...

Page 21: ...g System and are certified to serve as instructors in blood glucose monitoring using the EVENCARE G3 Blood Glucose Monitoring System They are qualified to perform initial training of certified operati...

Page 22: ...2 Use lancet lancing device correctly Yes No 3 Insert EVENCARE G3 test strip into EVENCARE G3 meter correctly Yes No 4 Apply blood to test strip correctly Yes No 5 Dispose of lancet correctly Yes No...

Page 23: ...23 IN SERVICE GUIDE EVENCARE G3 BLOOD GLUCOSE MONITORING SYSTEM REVIEW OF NURSE S TECHNIQUE DATE CERTIFIED TRAINER CERTIFIED TRAINER S SIGNATURE NURSE REVIEWED NURSE S SIGNATURE...

Page 24: ...EVENCARE G3 QUALITY CONTROL LOG DATE TIME STAFF PERFORMING TEST STRIP LOT LOW CONTROL HIGH CONTROL ACTION TAKEN LOT RANGE RESULT LOT RANGE RESULT...

Page 25: ...trol solution mode for low or high The ctl icon will display in the lower left corner of the LCD indicating the meter is in control mode STEP 1 STEP 2 STEP 5 STEP 3 STEP 4 STEP 6 STEP 1 STEP 2 STEP 5...

Page 26: ...55 128 Product is not made with Natural Rubber Latex Control Solution Range mg dL Reagent Composition Each test strip contains per cm2 of reaction zone approximately Glucose Oxidase A niger 2 3 Electr...

Page 27: ...es obtained from EVENCARE G3 Blood Glucose Test Strips are calibrated to plasma equivalents Blood glucose values will vary depending on food intake medication health stress and exercise The ideal rang...

Page 28: ...ing instead Blood Glucose Test Strips EVENCARE G3 EXPECTED VALUES Blood glucose values obtained from EVENCARE G3 Blood Glucose Test Strips are calibrated to plasma equivalents Blood glucose values wil...

Page 29: ...59 F 86 F DO NOT FREEZE OR REFRIGERATE DO NOT use past the expiration date Discard any unused control solution 90 days after first opening or after expiration date whichever comes first CONTROL SOLUT...

Page 30: ...abetes Care 2010 33 Suppl 1 S11 S61 Some products may not be available for sale in Mexico or Canada We reserve the right to correct any errors that may occur within this brochure 2017 Medline Industri...

Reviews: