Page 2: ...sponsibility however for possible errors or omissions or for any consequences resulting from the use of the information contained herein MEDIZS reserves the right to make changes in its products or pr...

Page 3: ...de 18 7 2 KER mode Keratometry mode 21 7 3 RK mode Continuous Keratometry and Refractometry mode 22 7 4 PK mode Peripheral Keratometry mode 23 7 5 CLBC mode Contact Lens Base Curve mode 27 7 6 IOL mea...

Page 4: ...tures of RK11 are as follows 1 RK11 offers various measurement modes 2 Because the RK11 covers a wide measurement range from 30D to 22D even an examinee with strong myopia can be measured 3 RK11 provi...

Page 5: ...he mains On power connection to the mains Do not throw away the waste to inappropriate place Manufacturer Authorized Representative in the European community Consult operating instructions 2 2 EU Coun...

Page 6: ...with the power supply indicated on the rating plate Otherwise it may result in fire or electric shock WARNING Be sure to turn OFF the power switch before connecting or disconnecting the cables Also d...

Page 7: ...ass in the monitor screen and on optical parts inside the instrument In this case just wait until condensation disappears before performing measurement 3 This instrument is used with accessories from...

Page 8: ...place If the inside temperate of instrument is too low it may cause on ERROR or inaccurate measurement 2 A sudden heating of the room in cold areas will cause condensation on the protective glass in...

Page 9: ...loth carefully to avoid any scratch 3 When moving this instrument please keep vibration or impact away from the equipment It can bring some damage inside or outside of the instrument 3 4 Storage space...

Page 10: ...ting height of Body Operation Lever It adjust the main body forward backward right side and left side for adjusting the focus Chinrest Up Down Buttons Regulating height of chin rest Power Switch Switc...

Page 11: ...ng the cables Also do not handle them with wet hands Otherwise you may get an electric shock that may result in death or serious injury 4 2 Back side of body Fig 2 Back side of body Name Description H...

Page 12: ...Stage Clamping Bolt Makes the system stage fixed Power Switch Switch for turning power ON and OFF Power IN LET Power IN LET RS 232 Connector Connect with external equipment EXT Video Connect with ext...

Page 13: ...IOL mode ON OFF RIGHT Auto Manual Select auto manual measurement UP Result Display measurement result DOWN Print Print measurement result WHEEL_LEFT Mode Select Select measurement mode Rotate the whe...

Page 14: ...e and do the same on left side Fig 7 Check whether stage is moving freely 5 2 Connecting the power cable Connecting the power cable to the connector on the bottom of main body Plug into AC outlet afte...

Page 15: ...he power 6 2 Set the model eye on Align the holes on the base of the model eye with the holes on the chin rest and then insert pins 6 3 Release stage lock Rotate the stage holding lever to UNFIXING di...

Page 16: ...heel up or down aims to the center RK11 does height auto tracking Operation the Body wheel Turning up down the Body Wheel the fine tuning height of main body Turning up down by pressing the Body Wheel...

Page 17: ...age is displayed repeat the procedure 6 5Adjust the height and focus on the model eye and measure again Check whether Diopter value is correct or not Diopter value is marked on the bottom of the model...

Page 18: ...former measurement results will be cleared automatically when staring next measurement NOTE Record on thermal paper can be spoiled by heat And Printed data on paper will be disappeared as time goes b...

Page 19: ...measurement UP Result Display measurement result DOWN Print Print measurement result WHEEL_LEFT Mode Select Select measurement mode WHEEL_RIGHT Mode Select Select measurement mode 1 Rotate the Operat...

Page 20: ...procedure RK11 analyze the focusing condition and display the level of focus Please refer to 6 5 5 Measurement Manual measurement mode MANU mode Press the Measure Button After finishing measurement r...

Page 21: ...layed maximum 10 results in order 6 Measuring another eye After finishing measurement for 1 eye move the lever to opposite side And then repeat the procedure 4 6 CAUTION When moving the stage to measu...

Page 22: ...urement result DOWN Print Print measurement result WHEEL_LEFT Mode Select Select measurement mode WHEEL_RIGHT Mode Select Select measurement mode 1 Rotate the Operation Buttons Wheel until KER mode is...

Page 23: ...and then measurement data will be printed 7 3 RK mode RK mode performs Refractometry and Keratometry at the same time Operation Buttons Wheel functions in RK mode KEY Function Description SET OK Confi...

Page 24: ...er required options were set properly 4 Adjust the height of examinee s eye and then perform measurement Refer to the section 6 5 and 7 1 3 7 5 Displaying measurement result Press Result UP button and...

Page 25: ...is displayed on the screen 2 Check whether PK mode is displayed on the screen 3 Measuring point mark on the screen will indicate the center 4 Focus to match with the height of examinee s eye Refer to...

Page 26: ...nter of cornea to the nose of examinee Temple TEM From the center of cornea to the temple of examinee Measurement is performed as below sequence SUP INF TEM NAS After measuring center of cornea periph...

Page 27: ...t result exists 8 Measuring another eye Move the stage to opposite side and then repeat the procedure 4 7 9 Displaying measurement result Press Result UP button then measurement result for periphery a...

Page 28: ...screen DOWN Print Print measurement result WHEEL_LEFT Mode Select Select measurement mode WHEEL_RIGHT Mode Select Select measurement mode 1 Rotate the Operation Buttons Wheel until CLBC mode is displa...

Page 29: ...e Measure Button 6 After measurement was finished result will be displayed on the screen Axis of base curve will be reverse of corneal axis 7 Displaying measurement result Press the Result Up button t...

Page 30: ...red with turning on IOL it can make less errors 1 Mode selection Rotate the Operation Buttons Wheel until REF or RK mode is displayed Press IOL LEFT button Check IOL icon is turned up 2 Shooting Keep...

Page 31: ...IGHT Mode Select Select measurement mode 1 Rotate the Operation Buttons Wheel until SIZE mode is displayed 2 Check whether SIZE mode is displayed on the screen 3 Adjust examinee s eye position by usin...

Page 32: ...e on right side of screen are displayed on the screen Rulers consist of 2 left right side ruler Press Select ruler SET OK of Operation Buttons Wheel then the mode will be selected 1st procedure is mov...

Page 33: ...odes 8 1 Result mode Seeing measurement result screen Show saved measurement result on LCD screen Measurement mode is changed to Result mode once Result UP button of the Operation Buttons Wheel was pr...

Page 34: ...lect Select the page RIGHT Page Select Select the page UP Item select Select the item DOWN Item select Select the item WHEEL_LEFT Value change Change the value WHEEL_RIGHT Value change Change the valu...

Page 35: ...printing function Eye Image Set print image option print image of emmetropia hyperopia myopia Screen Saver Set timer to power saving mode If RK11 is not operated until selected time power saving mode...

Page 36: ...measurement data units Kerato Unit Choose indication type of Keratometry RADIUS mm Corneal curvature DIOPT D Corneal refraction power Kerato Index Choose cornea equivalent refractive index Auto Start...

Page 37: ...24H Set Time AM 06 00 00 Brightness 0 10 Serial Baud Rate 9600 57600 115200 Factory Default Language English Spanish Korean Key Volume Choose the volume of sound Screen Saver Set timer to power saving...

Page 38: ...ET TIME Set current time Brightness Select the brightness of LCD screen Serial Baud Rate Choose serial communication speed with other equipment Factory Default Reload the factory data Initialize Confi...

Page 39: ...rning the number of printing ON Print the number of printing OFF Don t print the number of printing Print Type Choose printing format STANDARD Print all data AVERAGE Print average value only OFF Turn...

Page 40: ...on You can exit by pressing the SET OK button long time Operation Buttons Wheel functions in Set Message KEY Function Description LONG SET OK Save Exit Save Message And exit from Set Message SET OK Se...

Page 41: ...n 8 3 External Video Display RK11 supply RK image display function on external monitor Connect RK11 and external monitor with external video connector then user can check the image on external monitor...

Page 42: ...r switch and turn on again after 10 seconds If the message appears again consult the dealer CHART MOTOR ERROR IRIS SENSOR ERROR IRIS MOTOR ERROR 2 Message on Measuring Message Cause Remedy TRY AGAIN T...

Page 43: ...measurement can be performed in the IOL mode Examinee has IOL Intra Ocular Lens implanted Measure in the IOL mode 3 Message on printing Message Cause Remedy NO PAPER The printer is out of paper Replac...

Page 44: ...on right side of body As above figure insert the paper roll and then close printer cover 2 Replacement of Chinrest Paper Pull out 2 holding pins on chinrest Insert holding pins into the hole on chinr...

Page 45: ...ptom for fast service 2 Limit for supplying performance parts for repair Performance parts required to maintain the functioning of the product of the product will be stocked for six years after discon...

Page 46: ...iameter 2 0 12 00mm Storage Memory maximum 10 test results per L R eye Hardware Internal printer Thermal line printer Power saving Power saved after set time Go back once any button pressed 3 min 5 mi...

Page 47: ...46 12 Components Description Quantity 1 Power cable 1 2 Operation manual 1 3 Model eye 1 4 Printing paper 2 Rolls 5 Chin rest paper About 100 sheets 6 Dust cover 1 7 Fuse 2...

Page 48: ......

Page 49: ......