Medivators DSD-201 Site Requirements And Installation Download Page 1

DSD-201 and SSD-102 Site  
Requirements and Installation Instructions

 

50096-924 Rev A 

 

Page 1 of 12 

 
 
 

 

DSD-201 / SSD-102 
SITE REQUIREMENTS AND  
INSTALLATION 

Summary of Contents for DSD-201

Page 1: ...DSD 201 and SSD 102 Site Requirements and Installation Instructions 50096 924 Rev A Page 1 of 12 DSD 201 SSD 102 SITE REQUIREMENTS AND INSTALLATION ...

Page 2: ... a minimum of 1 2 providing a flow rate of 3 2 GPM with a maximum temperature of 110ºF 43ºC Cold water supply is preferred The incoming water pressure must be a minimum of 40 PSI and not to exceed 100 PSI A shut off valve must be installed before the pre filtration system It is important that the valve be mounted within the operator s reach 2 Drain The drain connection to the reprocessor is locate...

Page 3: ...rocessor 5 Space and Flooring The reprocessor contains leveling pads that can be adjusted during installation It is recommended to have a fairly level floor since the leveling ability of the included pads is limited It is required to have sufficient space to place the reprocessor along with all external components such as pre filtration water regulator water shut off valve and the ventilation syst...

Page 4: ...ation requirements have been met based on particulate analysis results NOTE this action is strongly recommended 1 Incoming water supply fed through a minimum 1 2 line 2 Intended drain outlet no more than 18 from the floor 2 Drain requirements met for the installation site 5 GPM 18 9 liters min 3 Fused outlet within 3 feet 1 meter of the installation site 4 Site adequately ventilated to meet OSHA r...

Page 5: ...ving dolly and ensure the reprocessor is not damaged The reprocessor must be installed on a level surface or be adjusted to level Adjust the leveling pads after unpacking the reprocessor During installation be sure the lower reservoirS cover is in place to prevent drill shavings and chips or other debris from entering the reservoirS 4 Water Supply Potable and Pre Filters If not already installed a...

Page 6: ...re regulator provides a flow rate of 3 2 gal min 12 liters min at a pressure 40 45 PSI 2 75 3 10 Bar Incoming water pressure to the regulator must be less than 90 PSI 6 2 Bar The external regulators water pressure is set between 40 45 PSI 2 75 3 10 Bar This is done during the flush cycle after the unit is ready for cycle verification ...

Page 7: ...DSD 201 and SSD 102 Site Requirements and Installation Instructions 50096 924 Rev A Page 7 of 12 12 50 inches Housing 2 Contains 0 45 micron filter Housing 1 Contains 1 micron filter ...

Page 8: ... between gauges on each side of a filter indicates a need for replacement For example if the pressure on gauge 2 is 35 PSI when the pressure on gauge 1 is 40 PSI a filter change is required 4 Installation Dimensions of the water filter are shown in figure 2 Note This unit is piped and assembled as shown with the outlet on the right side and gauge 1 incoming water pressure on the left figure 1 4 1 ...

Page 9: ...he drain tube must have 3 inches of drop 7 5 cm or greater over the 36 inch length Consult your maintenance department for multiple reprocessor installation Caution The reprocessor drain relies on gravity flow There must be no low spots or loops in the drain line Fluid trapped in the drain line will interfere with free drainage and reprocessor function 7 Electrical Caution The reprocessor does not...

Page 10: ...ectly Printer operates properly Reservoir heaters are set to the appropriate temperature required by the LCG manufacturer Basins fill and drain adequately during the disinfection phase within 90 seconds Basins fill and drain adequately during the rinse phase s within 90 seconds Alcohol and detergent reservoirs function properly level drops after each cycle Internal 0 2 µ filter installed and disin...

Page 11: ... the START button Check the flow through the tubing during the disinfect air and rinse cycle Check if incoming water pressure regulator provides a flow rate of 3 2 gal min 12 liters min at a pressure of 35 40 PSI 2 4 2 75 Bar Only adjust the External Water regulator during the Flush cycle Only adjust the internal regulators to 18 20 PSI during the Flush cycle Verify the cycle completes with no err...

Page 12: ...DSD 201 and SSD 102 Site Requirements and Installation Instructions 50096 924 Rev A Page 12 of 12 ...