2-8-633HAN/2-8-633CAS
HOOKUP CONNECTION GUIDE
Using Medivators
®
ADVANTAGE
®
Automated Endoscope Reprocessors
Port ID
Port Name
Connector Assembly Description
Order Number
1
LEAK TEST
Leak Test Connector Assembly
H28633-1
2
JET/AUX
Forward Water Jet Connector Assembly
H28633-2
3
BIOPSY
Biopsy Inlet Connector Assembly
H28633-3
4
AIR
Air/Water Connector Assembly
H28633-46
6
WATER
Air/Water Connector Assembly
H28633-46
7
SUCTION
Suction Connector Assembly
H28633-7
8
EXTRA
Biopsy2 Inlet Connector Assembly
H28633-8
A
NA
Air/Water & Suction Port Separator
1-5-313
B
NA
Selector Switch Cleaning Cap
1-5-381
If a Connector Assembly needs replacement, locate the port number/name and use
the corresponding order number found in the Connector Assembly Identification Chart.
Connector Assemblies are specific to individual hookups. Check each Hookup Connection
Guide for relevant information.
ADVANTAGE HOOKUP and CHANNEL SEPARATOR
(Figure 1)
* Representative endoscope for illustrative purposes only. For exact
port locations, refer to endoscope manufacturer’s instruction guide.
Each Advantage Hookup is comprised of multiple ports and connector assemblies. Each
port is labeled with a Port ID number (circled). This same number also identifies the
corresponding Connector Assembly. Each Connector Assembly is fitted with a specific
endoscope connector. (see Fig. 2).
EXAMPLE CONNECTOR ASSEMBLY
(Figure 2)
If you have any questions regarding this hookup, in the U.S., please contact Medivators Technical Support at
1-800-444-4729. Outside the U.S., please contact your Minntech Distributor.
READ ALL INSTRUCTIONS PRIOR TO CONNECTION
Locate the endoscope port* and follow the instructions for attaching the
corresponding Endoscope Connector. All connectors must be attached.
CONNECTOR ASSEMBLY IDENTIFICATION CHART
Ferrule
Endoscope
Connector
example
Tubing
Connector
Coupling
Label
Ensure proper alignment of all connectors with the
endoscope mating parts prior to attachment.
➀
➆
➃➅
A
➂
LOCK UNLOCK
Minntech Corporation
Minneapolis, MN 55447 USA
763.553.3300
Minntech BV
The Netherlands
+31.45.5.471.471
4
3
2
1
8
7
6
5
LEAKTEST
AI
R
AUX W
AT
ER
FW JET
ELEV
AT
OR
BIOPSY SUCTIO
N
JET/AUX
FW JET
AUX W
AT
ER
AIR
2
BALLOON GAS, BIOPS
Y
EXTRA
SUCTIO
N
WA
TE
R
BIOPSY
SUCTIO
N
➁
➇
B
Leak Test Connector
Align the slot on connector with
the pin on the scope. Push on and
turn clockwise to engage.
➀
Suction Connector
Align silicone tubing with
suction barb and push tubing
onto barb until fully engaged.
➆
Air/Water and Suction
Port Separator
Align the separator with the
piston ports and push to
compress spring before sliding
the frame forward to fully
engaged.
A
Biopsy Inlet Connector
Align stainless steel fitting
with biopsy port, push and
turn clockwise to connect.
➂
Air/Water Connector
Align air/water fitting with
endoscope port and insert.
Screw knurled collar clockwise
until fully engaged.
➃➅
Biopsy2 Inlet Connector
Align stainless steel fitting with
biopsy port, push and turn
clockwise to connect.
➇
Forward Water Jet
Connector
Align with forward water
jet port, insert and screw
clockwise until fully
engaged.
➁
Selector Switch
Cleaning Cap
Remove Channel Selector Switch.
Place cleaning cap over port and
insert. Press around edge until lip
is fully sealed.
B