23
Testing Control Samples
Prepare the test strip and Glucose Control
Solutions.
1.
Press the
button. The meter will
power on.
2.
Insert
MediSure
strip into the strip port.
3.
Press the
“>”
button to enter control mode. The
button “L1” refers to Level 1. You can select L1, L2 or
blood by pressing the
“>”
or
“<”
button.
4.
When the control bottle symbol
flashes on the
meter, it means that you can start a control test. You
have 3 minutes to apply the control solution. If you
do not apply the control solution within 3 minutes the
meter will turn off automatically.
5.
Use a clean, wet tissue to wipe the strip vial cap top.
Dry thoroughly.
6.
Roll Control Solution Bottle between hands to ensure
proper distribution of solution and open cap.
7.
Gently squeeze a drop onto the strip vial cap top.
8.
Let the tip of the test strip touch the control solution
drop.
9.
Allow the solution to be automatically drawn into
sampling window
until it is completely filled.
10.
After 6 seconds, the result will be shown on the
screen.
11.
Compare the control test result with the range
printed on the label of the test strip vial.