P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 21
-
Figure 4-4 Uninstalling Battery from Main Unit
4.3.3. Connecting and Disconnecting Probes
NOTE:
Ensure that the system is shut down before connecting and disconnecting probes.
Flip images horizontally to change the scan direction or vertically to change the image orientation.
The scan direction mark located at the side of
probe indicates the beginning direction of scanning.
The scan direction mark is shown below.
Figure 4-5 Probe Scan Direction Mark Schematic Diagram
There is information about Model and SN on the probe connector.
To connect a probe:
1.
Place the probe’s carrying case on a stable surface and open the case.
2.
Carefully remove the probe and unwrap the probe cable.
3.
Do not allow the probe head to hang free. Impact to the probe head could result in irreparable
damage.
Scan Direction Mark
①
Press the button
②
Pull the battery cover out
③
Pull the flicker counterclockwise to hide it
④
Pull out the battery