P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 135
-
Type
Linear array probe
Micro-convex probe:
Micro-convex probe:
Model
L761-2
C321-2
C611-2
B/M Frequency (MHz)
6.5/7.5/8.5
2.5/3.5/4.5
5.5/6.5/7.5
Harmonious
Frequency (MHz)
H9.0/H9.4
H5.0/H5.4
H9.0/H9.4
Doppler
Frequency
(MHz)
5.5/6.5
2.5/3.0
5.0/6.0
Needle Guide Bracket
Kit
BGK-LA70
BGK-CR20
BGK-MCR10
Application
Small
animal:
abdomen,
muscle,
tendon, small parts,
peripheral vascular
Small
animal:
obstetrics, abdomen,
cardiology.
Large animal: cardiology
Small animal: obstetrics,
abdomen, cardiology.
Large animal: cardiology
NOTE:
Maximum transducer temperature rise during use: Less than 10
C.
Expanded uncertainty of temperature test: U=0.4
C, k=2.
A1.7: Operating, Storage and Transportation Environment
A1.7.1. Operating Environment:
Please only use the device in the following environment:
Temperature
+5 °C ~ +40 °C
(+41ºF ~ +104ºF)
Relative humidity range
25% RH ~ 80% RH
(
no condensation
)
Atmospheric pressure range
860 hPa ~ 1060 hPa
A1.7.2. Storage and Transportation Environment:
Temperature
-20 °C ~ +55 °C(-4ºF ~ +131ºF)
Relative humidity range
25% RH ~ 93% RH
(
no condensation
)
Atmospheric pressure range
700 hPa ~ 1060 hPa