User Guide
MeCA812i
Warnings and Cautions
- Avoid humidity, high temperature, dust in the air, salt or sulfuric
material and avoid the unventilated place or the direct sunlight.
- Avoid vibration or mechanical shock.
- Avoid fire and explosion hazard. An explosion hazard may
result from the use of flammable anesthetics, skin cleansing
agents and disinfectants.
- Dispose of the packaging material, observing the applicable
waste control regulations. Keep the packaging material out of
children’s reach.
- If in doubt the integrity of the AC power source, the device
must be operated from its internal battery.
- Check the voltage rating before you connect the device to an
electrical outlet to ensure that the required voltage and frequency.
- If use multi-tap inevitably, use the one provided with earth pin
by all means.
- Unplug the AC power cord from AC outlet before the device
cleaning or disinfecting its surface.
- Do not spray, pour, or spill any liquid directly to the device for
cleaning or disinfection.
- When cleaning or disinfecting surface of the device or electrode,
fully dry and use it.
- Due to the risk of electrical shock, do not open the cover of
the equipment unless absolutely necessary and disassembly of
the machine should be done only by a maintenance worker who
has the license from MEDIGATE Inc.
Caution: Indicates a potential hazard.
If not avoided,
this hazard may result in minor personal injury and/or
product/property damage.
Warning: Indicates a hazard.
If not avoided,
the hazard can result death or serious injury.
- For your safety check the device before using.
- If put the electrodes too long time, they may cause skin problems.
Therefore, use them for measurement only
MEDIGATE Inc.
42-10, Taejanggongdan-gil, Wonju-si, Gangwon-do, Korea
Tel: (+82) 33-747-3552, Fax: (+82) 33-747-3691
© 2011-2014 MEDIGATE Inc. All rights reserved.
MeCA812i
User Guide
Part No-Rev. : L05351-03
Date of Revision: 2014-02-03
Printed in Korea
Default setting
1) Press the button to enter the Configuration menu.
2) Press Enter button on
System>Default Setting
.
Note:
After the default setting, all custom setting value
will be lost.
Checklist before using
□ Not physical damage on surface of device.
□ A Device was not get wet .
□ Rated power is supplied.
□ No physical damages on the patient cable.
□ No physical damages on the electrodes.
-Check the connection lead and electorde.
-Make sure all electrodes puts right place
and contacts well.
-If electrod gel gose dry it may block the contacts.
-Check the AC filter setting.
-Electrical blanket or heavy energy consume
electrical device may occur heavy noise.
Problem Check or Possible Solutions
You can adjust the Gain value
with the
Gain button
.
Recommended: 10mm/mV
You can adjust the Speed Value
with Speed button.
Recommended: 25mm/s
-Device needs some checks. See the operator’s
manual Chapter7.
-Use stronger Low Pass filter.
[
weak
150 - 100 - 40 (Hz)
strong
]
Note
: Stronger filter may make distortion.
-Enable the AFS. Refer to
Screen
-Use stronger High Pass filter.
[
weak
0.05- 0.16 - 0.3 - 0.5 (Hz)
strong
]
Note
: Stronger filter may make distortion.
-Too small or too big
ECG wave.
-ECG waves looks too
tight or too loose.
-There is too much
noise in ECG wave.
-I can’t recognize the
ECG signal.
-Error message shows
up after the operation
stops.
-High frequency noises are
added.
-Too much ECG baseline
wandering.
Troubleshooting
-Check the printer cover if it is opened.
-Check the printer cover may open.
-Grid side of paper must contact with
the printer header.
-Printer crumples up
paper.
-Printer does not work.
-Printer ejects blank
paper continously.
-Paper comes out but
not printed.
Medical Device
The electrocardiograph is intended to be used in a hospital and
hospital facilities environment by trained medical personnel.
For operator and patient’s safety, before you use the device,
read the
Operator’s manual
in enclosed CD.
This User Guide contains only minimum information of product.
Caution
Classification of electrical protection symbol:
This symbol
means the device is CF type(cardiac floating) protection applied.
Serial port symbol:
This symbol means Serial port (RS-232).
High Voltage Caution:
This symbol means high voltage parts included.
CE Marking:
CE Marking means this device is conformity
with provisions of the EU MDD(Medical device directive).
It is certified by DNV(NB No. 0434).
EU-Consult instructions for Use:
This symbol means read
manuals or descriptions from manufacture.
IPX0
IPX Level:
This symbol means waterproof level.
(IPX0 = No waterproof)
EU-Electronics and Battery Disposal Information:
This symbol means that the product and battery should
be recycled separately from household waste.
GND symbol:
This symbol means equivalent ground terminal
Network symbol:
This symbol means TCP/IP terminal
USB device symbol:
This symbol means Universal Serial Bus terminal.
SD card symbol:
This symbol indicates SD card slot.
EU-Warning or Caution:
This symbol indicates hazard.
If not avoided, the hazard can result death or serious injury,
minor personal injury or product damage.
Symbol Description
Symbols
Obelis s.a
Tel: (+32) 2-732-5954, Fax: (+32) 2-732-6003