Medica EasyLyte Na/K Service Manual Download Page 1

Cat. No. 2285 R3

EasyLyte

Service Manual

\EDIC@

Na/K
Na/K/Cl
Na/K/Li

Analyzers, Serial Numbers 5000 and above

EasySampler

Summary of Contents for EasyLyte Na/K

Page 1: ...Cat No 2285 R3 EasyLyte Service Manual EDIC Na K Na K Cl Na K Li Analyzers Serial Numbers 5000 and above EasySampler...

Page 2: ...IBRATE NOW ANALYZE BLOOD DAILY CLEANER STANDBY MODE SECOND MENU SEE RESULTS PACK USAGE QUAL CONTROL OPER FUNCTS DIAGNOSTICS DATE RANGES Whole Blood Serum or Plasma Analysis End of Day Cleaning Fluid C...

Page 3: ...Verification 23 Printer Verification 24 Sample Detector Verification 25 Signal Input Verification 26 Serial Communication Verification 28 QC Tests 29 Repair Repair Overview 33 Tools 34 Rear Housing Re...

Page 4: ...Housing Removal 91 Housing Installation 93 Spindle Column Removal 95 Spindle Column Installation 97 Sensor Removal 99 Sensor Installation 100 Motor Removal 101 Motor Installation 103 PC Board Removal...

Page 5: ...n and also serves as a waste receptacle for solution and samples after analysis System Overview The EasyLyte analyzer is engineered to make local service feasible Modular construction is used througho...

Page 6: ...l KEY INFORMATION Could take the EasyLyte analyzer out of operation WARNING Possible biological hazard to the operator EASYLYTE ANALYZER S Information for EasyLyte analyzers with serial num bers BELOW...

Page 7: ...ble connects the display to the CPU assembly Sample Detector Cable connects the sample detector to the display board Pump Assembly peristaltic pump to move fluids through the analyzer Pump Motor power...

Page 8: ...OSAMPLER P3 P1 P2 P4 P13 P6 Power Supply Cable Printer Cable Pump Motor Cable Sensor Cable Probe Motor Cable Display Cable Front Housing Display Assembly Sample Detector Cable Probe Position Sensor Pr...

Page 9: ...ct Description 5 Figure P2 Front Assembly Right Side View Front Housing Probe Motor Assembly Mounting Bracket Probe Position Scale Probe Motor Guide Rod Lead Screw Display Cable Printer Assembly Probe...

Page 10: ...6 Product Description Figure P3 Front Assembly Left Side View Front Housing Pump Head Assembly LCD Screen Pump Assembly Pump Motor Display Cable Printer Cable...

Page 11: ...line display Display Cable connects the display to the CPU assembly Sample Detector Cable connects the sample detector to the display board Pump Assembly peristaltic pump to move fluids through the an...

Page 12: ...wer Supply Assembly AUTOSAMPLER P3 P1 P2 P4 P13 P6 Power Supply Cable Printer Cable Pump Motor Cable Sensor Cable Probe Motor Cable Front Housing Display Assembly Sample Detector Cable Probe Position...

Page 13: ...P5 Front Assembly Right Side View Analyzers with Serial Numbers below 6100 Front Housing Probe Motor Assembly Mounting Bracket Probe Position Scale Probe Motor Guide Rod Lead Screw Printer Assembly V...

Page 14: ...Product Description Figure P6 Front Assembly Left Side View Analyzers with Serial Numbers below 6100 Front Housing Pump Head Assembly LCD Screen Pump Assembly Pump Motor Display Cable Printer Cable 61...

Page 15: ...ects presence of air liquid sets sample position inside electrodes Internal Filling Solution 2 molar KCI solution acts as salt bridge with reference electrode Fill Plug prevents the internal filling s...

Page 16: ...ply Figure P8 Rear Assembly Rear Assembly Components This section describes the components of the EasyLyte Rear Assembly Refer to Figure P8 for the location of each component Rear Housing rear case La...

Page 17: ...y from Medica CONSUMABLE ITEMS Description Catalog No Na Electrode 2102 K Electrode 2101 Cl Electrode 2113 Li Electrode 2106 Reference Electrode 2103 Sample Detector 2257 Sample Probe 2107 Solutions V...

Page 18: ...display 001937 006 EasyLyte Na K 1 line display 001937 007 EasyLyte Na K 2 line display 001937 008 EasyLyte Na K Chinese 001957 001 EasyLyte Na K Cl or Li Chinese 001957 002 EasyLyte Na K Japanese 00...

Page 19: ...nalyzer Since EasyLyte Analyzers with serial numbers below 6100 are assembled using Power Supply part no 05001065 and CPU part no 05001063 the troubleshooting instructions differ in certain areas Plea...

Page 20: ...Inspect EPROM Intallation Beeping or Power Malfunction DC Power Verification MTS Errors Probe Motor Probe Motor Verification Malfunction MTP Errors or Pump Motor Perform PUMP CAL test No Pump Movemen...

Page 21: ...end of the power cord Refer to Figure T1 For EasyLyte analyzers set to 110V the voltage at hot should be 90V 125V For EasyLyte analyzers set to 220V the voltage at hot should be 210V 240V If the volt...

Page 22: ...ect the EasyLyte entry module to a wall outlet with a known good power cord 8 Connect the reference side of the voltmeter to Pin 1 of the power supply cable 6 pin connector 9 Check the AC voltages at...

Page 23: ...power supply cable 13 pin connector from the CPU board see Figure T4 14 Connect the reference side of the voltmeter to Pin 5 of the powr supply cable Avoid shorting the adjacent pins 15 Check the DC v...

Page 24: ...below Test Point Value Tolerance TP1 5 00 V 0 15 V TP2 34 0 V 2 0 V TP3 5 00 V 0 15 V TP10 2 50 V 0 10 V TP11 5 00 V 0 15 V 5 For EasyLyte serial numbers below 6100 use TP11 as the reference point Ve...

Page 25: ...results take the appropriate corrective action Display Blank and YES NO buttons DO NOT operate Remove the test display cable Perform AC and DC power verification If the AC and DC power tests are succe...

Page 26: ...n connector as follows Remove the probe motor connector from the bottom left of the CPU Board position P3 Compress the strain relief cap against the probe motor connector to firmly seat the wires Rein...

Page 27: ...nnector into the CPU board 5 Repeat the Pump Test If the pump malfunctions replace the pump motor with a test motor See Pump Motor in the REPAIR SECTION for futher instructions 6 Repeat the Pump Test...

Page 28: ...in the printer head area Remove any obstructions if possible without damaging the printer assembly or paper tear bar If an obstruction cannot be removed replace the Printer Assembly Refer to Printer...

Page 29: ...detector installed indicates a bad connection Proceed to STEP 5 to check the sample detector circuit 5 Reinstall the sample detector Look at the left side of the sample detector A red light must shine...

Page 30: ...Mini Test High Impedance Voltage Source part no 001374 If this tool is not available the remaining tests cannot be performed 4 Connect the Voltage Source to the EasyLyte analyzer as follows Refer to F...

Page 31: ...cification Settings Group 1 Group 2 Polarity setting acid positive acid positive Impedance setting HI HI Offset setting 8 6pH 10 4 pH Voltage Values Individual values 59 mV 10mV 177 mV 10 mV Differenc...

Page 32: ...cifications Serial Interface Specifications Connector 25 pin D Sub male Pin type RS 232 Pin assignments Pin 1 Chassis Not Connected Pin 2 TxD Pin 3 RxD Pin 4 RTS Pin 5 CTS Pin 7 Signal Ground Transmis...

Page 33: ...meric value Numeric value wait 2 Two numbers appear on the display screen separated by an arrow A cursor will be posi tioned underneath the first digit of the first number The second number is not app...

Page 34: ...ONSTS Displays the factory set electrode constants The acceptable ranges are Left column number 0 92 to 1 08 Right column number 800 to 800 If values are out of specification contact Medica RAM PROM T...

Page 35: ...800 1200 Std B 800 1200 Std A 800 1200 Air 500 800 The three reagent values should be 100 200 of the Air value In most cases abnormal results indicate a faulty solutions valve NOTE The operator must p...

Page 36: ...ise noted Always unplug the power cord from the analyzer before beginning any repair work Note that only complete mechanical and electronic assemblies can be replaced Refer to the Replacement List in...

Page 37: ...r 1 Phillips Screw Driver 2 Allen Wrench 1 16 inch Scissors Hex Nut Driver 11 32 inch hollow stemmed Retaining Ring Pliers Resistor 10 OHM 5 watts Volt Meter Optional Tools The following is a list of...

Page 38: ...ry Module Power Cord Rear Housing Removal 1 Disconnect the power cord from the entry module see Figure R1 2 Detach the four screws from the rear housing using a 2 Phillips screw driver as shown in Fig...

Page 39: ...36 Repair Figure R3 3 Separate the front and back housing See Figure R3 Rear Housing Front Housing...

Page 40: ...ront and back housing See Figure R4 Verify that all cable connectors are attached properly Refer to Figures P1 and P6 in the PRODUCT DESCRIPTION SECTION for detail Verify that all wires are contained...

Page 41: ...ure R6 Figure R5 Entry Module Power Cord Screws Screws 2 Attach the four screws to the rear housing using a 2 Phillips screw driver as shown in Figure R5 3 Connect the power cord to the entry module s...

Page 42: ...gure R9 Figure R8 Figure R7 Power Supply Removal 1 Disconnect the power supply cable from the CPU board see Figure R7 Refer to Figure R8 for EasyLyte analyzers with serial numbers below 6100 2 Detach...

Page 43: ...Assembly Power Supply Assembly Figure R11 Figure R10 3 Lift and remove the power supply assembly from under the CPU board as shown in Figure R10 Refer to Figure R11 for EasyLyte analyzers with serial...

Page 44: ...e the power supply assembly under the CPU board as shown in Figure R12 Fit into the alignment holes in the bottom of the front housing Refer to Figure R13 for EasyLyte analyzers with serial numbers be...

Page 45: ...ctor to CPU Board Figure R16 Power Supply Cable Connector to CPU Board Figure R15 3 Connect the power supply cable to the CPU board as shown in Figure R15 Refer to Figure R16 for EasyLyte analyzers wi...

Page 46: ...3 Printer Bay Paper Roll Printer Lid Printer Figure R17 Figure R18 Printer Removal 1 Remove the printer lid from the printer see Figure R17 2 Remove the paper roll from the printer bay as shown in Fig...

Page 47: ...ter Cable Connector to CPU Board Printer Cable Connector to CPU Board Figure R19 3 Disconnect the printer cable from the CPU board see Figure R19 Refer to Figure R20 for EasyLyte analyzers with serial...

Page 48: ...ps Printer Cable Figure R22 Figure R21 4 Detach the printer cable from the retaining clamps located along the right side of the CPU board as shown in Figure R21 5 Detach the two screws using a 2 Phill...

Page 49: ...Printer Cable Printer Body Printer Body Figure R24 Figure R23 6 Remove the printer body as shown in Figure R23 7 Pull the printer cable up from behind the CPU board while holding the printer body as...

Page 50: ...nter Body Figure R27 Figure R26 Figure R25 Printer Installation 1 Slide the printer cable behind the CPU board as shown in Figure R25 2 Install the printer body into place See Figure R26 3 Attach the...

Page 51: ...ter cable to the CPU board see Figure R28 Refer to Figure R29 for EasyLyte analyzers with serial numbers below 6100 5 Using scissors cut the beginning of the paper roll into a V shape See Figure R30 I...

Page 52: ...ning Clamps Printer Printer Lid Figure R33 Figure R32 7 Place the printer lid onto the printer see Figure R32 8 Attach the printer cable to the retaining clamps located along the right side of the CPU...

Page 53: ...Display Cable Power Supply Cable Printer Cable Figure R36 Figure R35 Figure R34 CPU Board Removal 1 Disconnect the probe motor sensor pump motor power supply and printer cables from the CPU board See...

Page 54: ...ector See Figure R37 4 Disconnect the electrode jacks from behind the CPU board See Figure R38 for the location of the electrode jacks To disconnect gently pull electrode jack straight away from the c...

Page 55: ...r the tab as shown in Figure R39 6 Disengage the center plastic mounting post from the CPU board located above the serial and EasySampler ports 7 For EasyLyte analyzers with serial numbers below 6100...

Page 56: ...Repair 53 Mounting Posts Mounting Posts Figure R41 7 Slide the CPU board off the mounting posts See Figure R41...

Page 57: ...e R42 CPU Board Installation 1 Align the CPU board with the plastic mounting posts Snap the right side mounting posts of the CPU board into place See Figure R42 Do not snap the left or center mounting...

Page 58: ...proper connector pin position on the CPU board and the recommended installation order See Figure R44 for EasyLyte Analyzer with serial numbers below 6100 On some earlier EasyLyte analyzers below 6100...

Page 59: ...ble Display Cable Tabs Power Supply Cable Printer Cable Power Supply Cable Printer Cable Figure R48 Figure R47 4 Attach the display cable to the CPU board Push the display cable tabs to the front to s...

Page 60: ...Detach the two short self tapping screws from the sample detector board using a 2 Phillips screw driver See Figure R49 2 Move and hold the display cable to expose the three display assembly screws Det...

Page 61: ...soldered leads located on the top of the LCD screen the two line display has soldered leads located on the bottom of the LCD screen Refer to Figure R52 in the Display Installation for further details...

Page 62: ...so that the notch is located on the top inside edge of the adapter 2 Place the foam pad on the display board above the YES NO buttons 3 Place the LCD screen on top of the foam pad The orientation of...

Page 63: ...g so that the front of the analyzer is facing down Align the display assembly and insert into place as shown in Figure R53 using the YES NO buttons as a guide 5 Hold the display in place and attach th...

Page 64: ...r cable is positioned on the right side Attach the two screws lightly using a 2 Phillips screw driver see Figure R55 7 Install the sample detector in the front of the analyzer to verify the sample det...

Page 65: ...ump motor cable wires as shown in Figure R57 3 Detach the four screws from the front of the analyzer using a 1 Phillips screw driver See Figure R58 Some earlier EasyLyte analyzers may use socket screw...

Page 66: ...are positioned below the pump when properly aligned 2 Attach the four screws into the front of the analyzer using a 2 Phillips screw driver See Figure R61 Earlier EasyLyte analyzers may use socket scr...

Page 67: ...e Motor Assembly Removal 1 Detach the screw at the top of the probe position scale using a 2 Phillips screw driver as shown in Figure R62 Remove the washer Some EasyLyte analyzers use a white spacer t...

Page 68: ...haped retaining ring from the actuator using the retaining ring pliers DO NOT stretch out the retaining ring with the pliers Disengage the actuator from the top of the lead screw assembly See Figure R...

Page 69: ...R67 Figure R68 5 Detach the three screws from the probe motor bracket using a 2 Phillips screw driver as shown in Figure R67 Remove the washers The other two screws are located on the bottom edge of...

Page 70: ...robe Motor Assembly Installation 1 Place the probe motor assembly into place See Figure R69 2 Attach the three washers and screws to the probe motor assembly bracket using a 2 Phillips screw driver Se...

Page 71: ...e retaining ring onto the actuator using the retaining ring pliers DO NOT stretch out the retaining ring with the pliers 4 Place the probe motor cable wires into the retaining clamp See Figure R72 Sec...

Page 72: ...ned snap the probe into the probe arm B Manually move the probe arm until the sample probe hole is aligned with the top line marked on the alignment valve When the sample probe is aligned DO NOT move...

Page 73: ...ition Scale Figure R74 7 Hold the sample probe against the alignment valve tool to keep in place Install the probe position scale through the top of the housing see Figure R74 Verify the probe positio...

Page 74: ...tighten the screw Some EasyLyte analyzers use a white spacer that should be attached to the back surface of the scale 10 Align the probe position scale slot within the center of the probe position se...

Page 75: ...72 Repair Screw Figure R76 11 When alignment is complete fully tighten the probe position screw into place using a 2 Phillips screw driver as shown in Figure R76...

Page 76: ...e Arm Figure R77 Guide Rod Removal 1 Remove the sample probe from the probe arm and the solutions valve See Figure R77 Remove the solutions valve 2 Detach the screw from the bottom of the guide rod us...

Page 77: ...74 Repair Probe Arm Guide Rod Figure R79 3 Remove the guide rod and probe arm assembly down and away from the housing See Figure R79...

Page 78: ...stallation 1 Install the guide rod as shown in Figure R80 To install first insert the guide rod into the alignment hole located in the top wall of the front assembly 2 Attach the washer and screw to t...

Page 79: ...igure R84 Figure R83 Figure R82 Electrode Jack Removal 1 Insert the electrode jack wires into the 11 32 inch nut driver See Figure R82 2 Remove the hex nuts using the nut driver See Figure R83 3 Pull...

Page 80: ...ure R86 Figure R85 Figure R87 Electrode Jack Installation 1 Push the electrode jacks through the holes in the front of the analyzer see Figure R85 2 Insert the electrode jack wires into the 11 32 inch...

Page 81: ...nd The EasySampler contains a 25 position sample tray which includes a STAT and two quality con trol positions The tray holds either 500 l or 2 ml sample cups The EasyLyte analyzer software allows the...

Page 82: ...and receiver Chassis Assembly frame which supports the mechanical assemblies PC Board Assembly contains the PC board and IR transmitter and receiver Motor Assembly powers the spindle column assembly D...

Page 83: ...rchased directly from Medica Description Part No EasySampler Housing 001074 IR Windows 001192 Chassis Assembly 001370 PC Board Assembly 001233 Motor Assembly 001328 Spindle Column Assembly 001680 Sens...

Page 84: ...m Source Corrective Action Steps LCD screen displays SAFETY BEAM Verify there is no IR obstruction OBSTRUCTION and OBSTRUCTION Sensor Verification analyzer is beeping LCD screen displays MTS INCORRECT...

Page 85: ...screen press YES CALIBRATE NOW appears on the LCD screen press YES Return to normal operation Probe Misaligned If the probe is misaligned proceed to STEP 5 5 Grasp the EasySampler and analyzer at the...

Page 86: ...t cup position 12 3 CLEANER IN TRAY appears on the LCD screen press NO 4 The sample tray should rotate and then stop with the probe at Daily Cleaner position 5 Based on the test results take the appro...

Page 87: ...tray should spin one revolution then stop 4 FIND HOME appears on the LCD screen press YES The sample tray should spin and stop with the probe at the Daily Cleaner position 5 Based on the CALIB CAROUS...

Page 88: ...ily Cleaner position to each cup position and back for cup positions 11 through STAT 5 The sample tray will perform four complete rotation cycles as described above 6 The analyzer will print the test...

Page 89: ...Manually turn the sample tray off the Daily Cleaner position 6 Verify that the measured voltage is less than 0 50 volts 7 Based on the voltage test results take the appropriate corrective action In Sp...

Page 90: ...umeric value Numeric value wait 2 Two numbers appear on the screen separated by an arrow A cursor will be positioned underneath the first digit of the first number The second number is not applicable...

Page 91: ...OUSEL Verifies EasySampler movement functions CALIB CAROUSEL Moves sample tray one revolution FIND HOME Returns sample tray to Daily Cleaner position CYCLE TRAY Performs four complete tray cycles to e...

Page 92: ...to the Replacement List in this section which lists replaceable assemblies and appropriate catalog numbers Recommended Tools The following is a list of the tools required to perform repairs on the Eas...

Page 93: ...emoval 1 Turn the EasySampler upside down and gently place on a flat surface 2 Remove the two screws from the cable retaining plate using a 2 Phillips screw driver as shown in Figure A2 Figure A2 Scre...

Page 94: ...e EasySampler cable from the groove in the bottom of the housing 4 Remove the three screws from the metal base using a 2 Phillips screw driver as shown in Figure A3 5 Turn the EasySampler to the uprig...

Page 95: ...lign and place the housing over the chassis assembly as shown in Figure A5 Be careful not to damage the IR receiver or transmitter on the PCB 2 Turn the EasySampler upside down and gently place on a f...

Page 96: ...mpler cable into the groove in the bottom of the housing See Figure A7 5 Attach the two screws to the cable retaining plate using a 2 Phillips screw driver as shown in Figure A7 Screws EasySampler Cab...

Page 97: ...lumn Removal 1 Disconnect the sensor cable and the motor cables from the PC Board See Figure A8 2 Turn the adjustment knob counterclockwise to loosen Remove the adjustment knob from the bottom of the...

Page 98: ...96 EasySampler Spindle Column Assembly Figure A10 3 Remove the spindle column assembly as shown in Figure A10...

Page 99: ...A11 Spindle Column Installation 1 Place the spindle column assembly down into place as shown in Figure A11 2 Attach the adjustment knob see Figure A12 and turn clockwise to tighten Some EasySamplers...

Page 100: ...98 EasySampler Motor Connector J1 Sensor Connector J3 Sensor Cable Motor Cable Figure A13 3 Attach the sensor cable and the motor cable to positions J1 and J3 on the PC board see Figure A13...

Page 101: ...Loosen the two slotted screws from the motor bracket using a 2 Phillips screw driver as shown in Figure A14 2 Remove the sensor and mounting block as shown in Figure A15 3 Remove the two slotted screw...

Page 102: ...ace the sensor into the mounting block 2 Attach the two slotted screws to the mounting block using a 2 Phillips screw driver 3 Place the sensor and mounting block into the motor bracket See Figure A16...

Page 103: ...mper Damper Socket Screw Figure A19 Figure A18 Motor Removal 1 Loosen the socket screw from the brass damper using a 1 16 inch Allen wrench See Figure A18 2 Remove the damper from the top of the motor...

Page 104: ...Shelf Motor Slotted Screws Motor Figure A21 Figure A20 3 Remove the four slotted screws from the top of the motor using a 7 32 flat head screw driv er See Figure A20 4 Remove the motor from the motor...

Page 105: ...Place the motor into the motor shelf 2 Pull the belt around the gear of the motor see Figure A22 for location of gear and belt 3 Attach the four slotted screws to the top of the motor using a 7 32 fla...

Page 106: ...104 EasySampler Socket Screw Damper Figure A24 5 Tighten the socket screw to the damper using a 1 16 inch Allen wrench See Figure A24...

Page 107: ...nk Plate Hex Nut Power Cable Figure A26 Figure A25 PC Board Removal 1 Disconnect the power cable from the PC board See Figure A25 2 Remove the hex nut from the heat sink plate using a 1 4 inch open en...

Page 108: ...e A27 3 Remove the four screws from the PC board using a 2 Phillips screw driver See Figure A27 The lower left screw may be hidden by the EMI suppressor box If so peel back the tape to reveal the scre...

Page 109: ...Screws Screws Figure A30 Figure A29 PC Board Installation 1 Place the PC Board as shown in Figure A29 onto the base housing 2 Install the four screws to the CPU board using a 2 Phillips screw driver...

Page 110: ...nnector J2 Heat Sink Plate Hex Nut Power Cable Figure A32 Figure A31 3 Install the hex nut to the heat sink plate using a 1 4 inch open end wrench see Figure A31 4 Connect the power cable to position...

Page 111: ...n n 20 C V 1 C V 2 SD 03 mmol L C V 2 between run 10 day C V 2 C V 2 5 SD 05 mmol L C V 2 5 Accuracy typical Flame Photometer Amperometric correlation coeff 0 99 1 00 0 99 0 97 slope 0 99 1 07 1 09 0...

Page 112: ...use settings Size Weight 9 5 W x 16 5 H x 8 0 D 13 lbs 24 cm x 42 cm x 20 cm 5 8 kg EasySampler Specifications Size Width 12 2 inches 31 0 cm Depth 14 0 inches 35 6 cm Height 8 8 inches 22 4 cm Weight...

Page 113: ...rectly AIR IN STD B Standard B not detected correctly AIR IN FLUSH Flush not detected correctly ALL ABNORMAL Allows operator to delete all stored abnormal control results ALL NORMAL Allows operator to...

Page 114: ...NGE PACK Occurs at 99 pack usage Pack must be replaced CLEANING Analyzer is performing daily cleaning Cl HIGH Chloride result obtained is higher than the preset normal range flashing display Cl LOW Ch...

Page 115: ...MARY Allows operator to display summary of all results in storage ERROR Denotes an error has occurred Press YES or NO to determine error FLOW TEST FAIL Unsuccessful Fluid Flow Test FLUID FLOW OK Succe...

Page 116: ...Potassium electrode voltage is outside operating limits MV RANGE Li Measured Lithium electrode voltage is outside operating limits MV RANGE Na Measured Sodium electrode voltage is outside operating li...

Page 117: ...lows operator to print all results stored in memory PRINT STAT TICS Allows operator to print normal or abnormal control summaries PRINT SUMMARY Prints daily summary of results PRINT WKLY SUM Prints su...

Page 118: ...is mode SAMPLE CAL Analyzer is calibrating the Sample Detector SAMPLE DETECT OK Successful completion of Sample Detector Test SAMPLE DETECTOR Sample Detector readings out of range SAMPLE DET TEST Allo...

Page 119: ...IN Allows operator to perform additional purge after initial failure URINE Cl LIMITS Allows operator to set normal Cl range for urine URINE K LIMITS Allows operator to set normal K range for urine URI...

Page 120: ...E CALIBRATE NOW RUN CONTROLS DIAGNOSTICS TEST VALVE TEST ALIGNMENT BURN IN QC TEST CAL VALUES ELECTRODES VALUES FLUID FLOW TEST SAMPLE DET TEST EASYSAMPLER 5 NORMAL AND 5 ABNORMAL CONTROLS AT CUP 1 TO...

Page 121: ...CPU Board Analyzer Series below 6100...

Page 122: ......

Page 123: ...CPU Board...

Page 124: ......

Page 125: ...Power Supply Analyzer Series under 6100...

Page 126: ...Transformer Assembly...

Page 127: ...EasySampler PCB...

Page 128: ......

Page 129: ......

Page 130: ...001462 001 R3 EDIC Medica Corporation 5 Oak Park Drive Bedford MA 01730 For Service Or Technical Assistance Call 1 800 777 5983 In Continental US 1 617 275 4892 International...