babyDOPPY
Ultrasonic fetal heartbeat detector
USER MANUAL
1011
v.: 1.
Page 1: ...babyDOPPY Ultrasonic fetal heartbeat detector USER MANUAL 1011 v 1...
Page 2: ...kiad s 1 Ultrasonic fetal heartbeat detector Medi CAD Kft 3529 Miskolc L z r Vilmos u 4 Tel 36 46 369 377 431 395 Fax 36 46 560 423 http www medicad hu...
Page 3: ...hecking and charging the battery 5 6 Cleaning and disinfection 7 7 Environmental and other conditions of the use storage and transportation of the device 8 8 Technical specifications 10 9 Contents of...
Page 4: ...g its 2 MHz sensor human fetal heart activity can be detected from the second trimester of pregnancy From then the fetal heart rate can be monitored to the end of pregnancy On the LCD panel located on...
Page 5: ...2 Main parts of the device Sensor Push button On Off switch Volume control wheel VOLUME Headphone socket PHONE Sensor battery charger connector PROBE LCD display Charger adapter 3...
Page 6: ...nsor and move it slowly along the mother s abdomen until you can detect a clear distinctive fetal heart sound You can achieve the best quality heart sound when the sensor head is in the most ideal pos...
Page 7: ...the sensor connector and pull out the connector The volume can be adjusted with the VOLUME control wheel Attention The volume control wheel is not the on off switch the wheel serves only to adjust th...
Page 8: ...insert the plug of the network charger supplied by the manufacturer into the socket with the PROBE sign in the socket where the sensor was previously connected and insert the plug of the charger into...
Page 9: ...d the sensor Never submerge the sensor or the main unit in any liquid e g water alcohol oil as they may be irreparably damaged Attempt to sterilize the device with hot water or steam may result in the...
Page 10: ...evice must not be exposed to direct sunlight for more than 10 minutes The device must be protected during use transportation and storage from impact and other excessive mechanical stress e g dropping...
Page 11: ...At the end of its life cycle the device must be disposed of at the manufacturer or at collection sites specialized for the recycling and disposal of electronic equipment 9...
Page 12: ...heart rate FHR calculation Calculated based on the average of 8 heartbeats Range 60 min 210 min Classification Device operated by internal power source BF type patient side part Power supply 9V recha...
Page 13: ...with a fully charged battery Dimensions main unit 117 x 78 x 25 mm sensor 16 20 x 105 mm Weight main unit approx 150 g sensors approx 120 140 g each charger approx 70 g The manufacturer reserves the...
Page 14: ...9 Contents of the kit babyDOPY unit 1 pc Sensor 1 pc Conductive gel 1 tube Charger adapter 1 pc Case 1 pc User s manual 1 pc 12...
Page 15: ...nance The repair and maintenance of the device may only be performed by Medi CAD Kft or its authorized technicians Address Medi CAD Kft 3529 Miskolc L z r Vilmos u 4 Tel 36 46 369 377 431 395 Email me...
Page 16: ...cturer Medi CAD Kft H 3529 Miskolc L z r Vilmos u 4 Hungary Tel 36 46 369 377 Fax 36 46 560 423 Product Ultrasonic fetal heartbeat detector Type DOPPY product family babyDOPPY Classification Class II...
Page 17: ...1 _________________________ This warranty is valid for the above product for 24 months from purchase In case of malfunction the product is repaired free of charge within the warranty period Main place...