Hardware Description
23
amplification circuit are used to detect the patient side occlusion pressure during infusion, and the EEPROM
chip is used to record the initial value of the pressure sensor.
b.
Main test point: None
c.
Circuit board socket numbering and definition
No.
Socket
PIN
Quantity
Name
Description
1
J1
16
Socket for connection to J11
of HP-60 main board
/
2
J2
3
Socket for connection to
HP-60 door potentiometer
assembly
/
3
J3
10
Tube detection board and
anti-free-flow clamp board
signal terminal
Used for connection to J1 of the
HP-60 tube detection board and J1
of the HP-60 anti-free-flow clamp
board.
4.10
HP-60 Tube Detection Board
a.
Circuit description:
(1) The tube detection board contains a Hall switch assembly, which is used to detect whether the pump
door is closed. The pump door contains a magnet. When the pump door approaches, the signal of the Hall
switch assembly changes. Whether the door is closed is judged by detecting signal changes.
(2) The fluid side occlusion pressure detection board contains an optical coupling circuit, which is used to
detect the infusion tube. When an infusion tube is installed, the light baffle shields the optical coupler,
indicating that an infusion tube is installed.
b.
Main test point: None
c.
Circuit board socket numbering and definition
No.
Socket
PIN
Quantity
Name
Description
1
J1
5
Socket for connection to J3
of
HP-60
patient
side
occlusion pressure detection
board
/
4.11
HP-60 Anti-Free-Flow Clamp Board
a.
Circuit description:
The anti-free-flow clamp board contains an optical coupling circuit, which is used to detect the
anti-free-flow clamp signals of the simple dedicated infusion set. When the anti-free-flow clamp is installed,
the light baffle shields the optical coupler.
b.
Main test point: None
c.
Circuit board socket numbering and definition
No.
Socket
PIN
Quantity
Name
Description
1
J1
3
Socket for connection to J3
of
HP-60
patient
side
occlusion pressure detection
board
/
4.12
Stepper Motor
a.
HP-60 uses two types of stepper motor, that is, door opening motor and infusion motor. The door opening
motor connects to the main board J8 to control door opening and closing. The drive voltage of the infusion
motor is 24V, 2-phase drive.
b.
Socket numbering and definition: