|  CONNECTIONS 

The lever extension motor is activated with a single output control line.  By turning the output 

bit “ON” the lever will extend into the chamber.  When the output bit is turned “OFF” the lever 

will retract out of the chamber.  When the animal depresses the lever the signal is received on 

the input bit. 

Operate Connector 

Operate connects to any available Output on a Standard MED Connection Panel via an included 

SG-216A cable. 

Molex Pin # 

Function 

Wire Color 

28 Volt Ground 

Black 

Extend/Retract Lever 

White 

+28 Volts 

Red 

Response Connector 

Response connects to any available Input on a Standard MED Connection Panel via an included 

SG-216A cable. 

Molex Pin # 

Function 

Wire Color 

28Volt Ground 

Black 

Lever Response 

White 

+ 28 Volts 

Red 

*Optional  Lever Position Connector 

A lever position feature has been added to the circuitry of this low profile lever.  When the lever 

is extended, a signal is sent to an Input bit.  The signal is turned off when the lever is retracted. 
Lever Position connects to any available Input on a Standard MED Connection Panel via an 

included SG-216A cable. 

Molex Pin # 

Function 

Wire Color 

28Volt Ground 

Black 

Lever Position 

White 

+ 28 Volts 

Red 

  
 

Summary of Contents for ENV- 112CM

Page 1: ...for research RETRACTABLE RESPONSE LEVER ENV 112CM PRODUCT MANUAL DOC 050 Rev 1 4 Copyright 2017 All Rights Reserved Med Associates Inc P O Box 319 St Albans Vermont 05478 Phone 802 527 2343 Fax 802 52...

Page 2: ...i DOC 050 Rev 1 4 Copyright 2017 Med Associates Inc notes...

Page 3: ...apter 1 Introduction 1 Specs 1 Chapter 2 Connections 2 Operate Connector 2 Response Connector 2 Optional Lever Position Connector 2 Chapter 3 Adjustable Tensioner 3 Chapter 4 Maintenance 3 Cleaning 3...

Page 4: ...V 112CM permits placement just 2 1cm above the grid floor and can be raised as needed with chamber filler panel inserts This lever is designed to work with ENV 007 ENV 008 and ENV 009 series chambers...

Page 5: ...n Function Wire Color 1 28 Volt Ground Black 2 Extend Retract Lever White 3 28 Volts Red Response Connector Response connects to any available Input on a Standard MED Connection Panel via an included...

Page 6: ...done once a week or once a month based on use to reduce buildup of dander dust or debris which can lead to unreliable operation and response detection Humid environments may require more frequent clea...

Page 7: ...of lever 2 Using a 1 Philips screwdriver remove the 4 screws 2 on each side shown below 3 Slide the case backwards making sure not to pull completely off as there are wires connecting the circuit boar...

Page 8: ...4 again to blow out any remaining debris 7 Slide the case forward and reattach using the 4 screws removed in step 2 Reinstall lever in chamber APPENDIX A CONTACT INFORMATION Please contact Med Associ...

Reviews: