Waveform:
positive square wave
Amplitude:
> 1 V
pk
/ 50 Ω
Division ratio:
off, 1, 10, 100
Rise time:
< 1 ns
Output terminal:
coaxial N connector
Waveforms:
DC (positive, negative), square wave (symmetrical, positive, negative)
Frequency range:
0.100000 Hz – 100.0000 kHz
Frequency uncertainty:
25 ppm
Duty cycle range:
1.00 – 99.00 %
Amplitude range:
DC:
0.000 mV – 10.500 V
Square wave: 1.000 mV
pk
– 10.500 V
pk
Amplitude uncertainty:
0.1 % + 50 µV at 10 kHz and below
0.2 % + 50 µV above 10 kHz
Max. load current:
20 mA (1 V
pk
at 50Ω load)
Rise time:
< 200 ns
Output terminal:
4mm terminals HI – LO
Waveforms:
DC (positive, negative), symmetrical square wave
Frequency range:
15.00000 – 1000.000 Hz
Frequency uncertainty:
25 ppm
Amplitude range:
DC:
0.000 mV – 200.00 V
Square wave:
1.000 mV
pk
– 200.00 V
pk
Amplitude uncertainty:
DC:
as per DCV specifications
Square wave:
0.3 % + 50 µV
Max. load current:
as per ACV specifications