Multimeter function has direct access to several subfunctions. Push SELECT key until active window
Meter is selected. Meaning of soft keys beneath the display is changed. Soft keys enables fast access to
following functions:
-
Smoothing
Smoothing filter activation
-
Scaling
Allow to scale measurements relative to a reference value
-
Statistics
Statistical data can be displayed
-
Limits
Limits of band setting
-
Range
Auto or fix range selection
The function enables activation of smoothing filter and setting integration time in terms of number of
samples which are averaged. Set item Smoothing to ON to activate the filter.
The Scaling function allows recalculation of multimeter readings. Press Scaling softkey to set scaling
parameters:
-
Scaling. Use this item to turn scaling ON. Scaling is automatically turned off after each meter
function change (for example changing from DCV to ACV).
-
Function. „%“ shows relative value and „Scale“ shows absolute value, based on formulas
and
.
-
Gain, Offset and % Reference. Scaling formula parameters.
-
Custom unit and Custom unit label. Custom unit with length up to 4 alphanumerical characters can
be defined. Use navigation buttons or rotary knob to enter requested letters, switch between
uppercase and lowercase using A ↔ a softkey and enable this feature by turning Custom unit ON.
Rel. value [%] = (reading - Reference) / Reference ∙100
Abs. value = Gain ∙ r Offset
Summary of Contents for 9010+
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 33: ...Following items can be modified ...
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......
Page 55: ......
Page 57: ......
Page 58: ......
Page 78: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual ...