Low resistance mode is available in two- terminal connection only with maximum test from 2 to 5 V
depending on set value. See specification for test voltage and test current limitations.
Push MODE soft key to display list of modes. Select Variable and confirm with ENTER.
1.
Connect the multimeter to the calibrator as shown below.
2.
Select appropriate function and range on the multimeter.
3.
Push CAPACITANCE function key on the calibrator.
4.
Switch output terminals on by pushing key OPER.
5.
Push STBY key to switch output off.
Warning: If you are using RCL meter with 4TP connection or grounded equipment, proceed with
caution. Due to using simulation principle it is necessary to swap Hi and Lo terminal. High terminal of
RCL must be inserted into Lo terminal of calibrator and Lo terminal of RCL must be inserted into Hi
terminal of calibrator. Do not try to solve this issue with isolation transformer otherwise accuracy of
calibrator will be disturbed.
Fixed mode offers high accuracy of sourced capacitors with fix nominal values.
Push MODE soft key to display list of modes. Select Fixed item and confirm with ENTER. Selection of
individual fix value positions can be done using softkeys
, or cursor keys or using rotary button.
Numerical keypad is not active in this mode.
Capacitance SETUP can be activated using SETUP SOURCE key. Following parameters are available:
-
Terminal ground
LO output terminals are internally connected to PE wire when parameter is set to ON.
Calibrator can supply output voltage and current at the same time. The function is designed for
calibration of electric power meters and power analyzers. Push function key POWER to select
Power/energy mode.
Terminals
HI
LO
I+
I-
HIR
GND
Entry
Function
TRG
HF
9010
Multifunction Calibrator
V
COM
A
R
METER
SCOPE
SOURCE
METER
NULL
SELECT
SETUP
SYSTEM
1100V
RMS
MAX
1500V
DC
MAX
1V PK
MAX
20V PK
MAX
1 00V PK
MAX
10V PK
MAX
2V PK
MAX
25V PK
MAX
10V PK
MAX
50V PK MAX
50 V PK MAX
FUNC
OPER
STBY
10V
RMS
MAX
Summary of Contents for 9010+
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 5: ......
Page 33: ...Following items can be modified ...
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 50: ......
Page 51: ......
Page 55: ......
Page 57: ......
Page 58: ......
Page 78: ...Temperature range 250 00 1820 00 C Cold junction comp modes Manual ...