Harmonic preset feature works similarly as normal preset feature. Harmonic preset allows you to save
and recall one of up to 22 configurations. First preset is 00 Sine and it can’t be changed or saved.
If the preset is being edited and not saved, an * appears in front of the name on the main panel.
Harmonic preset is selected either from the main screen of the VAC, IAC, PAC functions from
"Harmonic" drop-down menu or from the HARMONIC PRESET screen after activating the LOAD key.
20 mV and 200 mV DC ranges can operate in either passive or active output mode, all other ranges
operate in active output mode only.
Active output is formed by electronic amplifier. The output offers higher load current up to 5 mA
however with a little bit worse uncertainty. Passive mode is formed by resistance divider with output
resistance 50 Ω. Output signal is less noisy and with better uncertainty however maximal output current
is given by its output impedance. See specification for detailed differences.
Push SELECT key until field Output mode is not activated. Change setting from Active to Passive using
soft key arrow keys, soft key List or rotary knob.