Se ng up the WebDAQ

Connec ng to a local area network

The WebDAQ 504 requires a TCP/IP connec on to a network or the Ethernet port of a computer. A standard

Ethernet cable is shipped with the device. The WebDAQ IP address type is set by default for DHCP/Link-Local.

When connected to a network, the device first a empts to receive an IP address from a DHCP server. If this

fails, a link-local address is used. Use of a sta c address requires configura on via the web interface. The IP

address iden fies the WebDAQ 504 on the network, and is used to access the web interface.

Connec ng to a network with DHCP enabled

Connect one end of the Ethernet cable to the connector on the WebDAQ rear panel, and connect the other end

to a 10Base-T or 100Base-TX compa ble Ethernet port, hub, or switch.
Once the device is detected, DHCP assigns an address. A different address may be assigned each  me the

hardware is connected to the network.

WiFi support

A er successfully configuring a wired connec on, you can use the web interface to configure the

WebDAQ for wireless communica on. Refer to the 

Configuring the WebDAQ for wireless

communica on

 sec on for more informa on.

Connec ng directly to a PC

Connect one end of the Ethernet cable to the connector on the WebDAQ rear panel, and connect the other end

directly to the Ethernet port on a PC. The WebDAQ a empts to communicate with the host PC using

169.254.100.100 – the link-local address stored on the device. The host PC may accept this address or use a

different address. The link-local address is valid only for communica ons between the WebDAQ and the host PC

to which it is connected.

Connec ng the external power adapter

Connect the network cable before connec ng to external power

Connec ng to the external supply before installing on a network results in the device a emp ng to

auto-nego ate a network address before one is available.

An external power adapter is included with each shipment (PS 9V1AEPS230V).

Connect the adapter to the power connector labeled EXT PWR on the device rear panel, and plug the AC

adapter into an electrical outlet.

Press the POWER bu on. The POWER LED on the rear panel turns yellow as the hardware starts up, and green

when it is ready for use.

WebDAQ 504 User's Guide

page 5

Summary of Contents for WebDAQ 504

Page 1: ...any form by any means electronic mechanical by photocopying recording or otherwise without the prior wri en permission of Measurement Compu ng Corpora on No ce Measurement Compu ng Corpora on does not...

Page 2: ...for reaching Tech Support For interna onal customers contact your local distributor Refer to the Interna onal Distributors sec on on our website at www mccdaq com Interna onal Hazardous voltages Take...

Page 3: ...The WebDAQ is designed with the Raspberry Pi Compute Module featuring a quad core processor with speeds up to 1 2 GHz 1 GB RAM and 4 GB Flash memory Integrated opera ng system and web service The Web...

Page 4: ...e too large error is returned and the acquisi on stops Before star ng a logging session of this type reformat the media using a di erent le system such as NTFS Refer to the Frequently Asked Ques ons h...

Page 5: ...e the WebDAQ for wireless communica on Refer to the Con guring the WebDAQ for wireless communica on sec on for more informa on Connec ng directly to a PC Connect one end of the Ethernet cable to the c...

Page 6: ...indow and enter h p webdaq xxxxxx local where xxxxxx is the last 6 digits of the MAC address The MAC address is on a label at the bo om of the WebDAQ enclosure If your network is con gured for the zer...

Page 7: ...a ng system and restart the WebDAQ using the Power bu on Con guring the network router for communica on across networks To communicate with the WebDAQ over the internet from a computer connected to a...

Page 8: ...ngs To restore network con gura on and security se ngs to the factory default values use a paper clip to press the CONFIG RESET bu on When pressed the default network and security se ngs are wri en to...

Page 9: ...ger an alarm and be driven high or low when an alarm occurs BNC connectors Connect up to four analog inputs to the BNC connectors AI0 AI0 to AI3 AI3 All analog inputs are simultaneously sampled at rat...

Page 10: ...connec ng a grounded signal source to the WebDAQ 504 Figure 6 Connec ng a grounded IEPE sensor Figure 7 shows a diagram for connec ng a oa ng signal source to the WebDAQ 504 Figure 7 Connec ng a oa ng...

Page 11: ...d to reboot the device release the bu on when the POWER LED blinks yellow Refer to Power On Mode for the procedure to remove jumper W1 and enable Power On mode The WebDAQ device can also be restarted...

Page 12: ...duled start condi on is met wai ng for job start condi on to be met Blinking green Data is being acquired job start condi on is met O No error detected 3 ERROR Blinking yellow Error condi on is detect...

Page 13: ...data to a single le when logging to external media do not use media forma ed with the FAT32 le system FAT32 is limited to a 4 GB maximum le size If the log le on FAT32 media grows beyond that size a F...

Page 14: ...ly Asked Ques ons help topic for more informa on about le systems Analog input circuitry The analog input channels are referenced to chassis ground through a 50 resistor To minimize ground noise make...

Page 15: ...nd signals The lters discriminate between signals based on the frequency range or bandwidth of the signal The three important bandwidths to consider are the passband the stopband and the alias free ba...

Page 16: ...quency Logging data You can log up to 3 GB of data to internal WebDAQ storage To generate larger data les log data to external media with a storage capacity suitable for the acquisi on USB drives and...

Page 17: ...round screw 4 Remove the nut and washer from the EXT PWR connector 5 Remove the two front panel bezel screws and remove the front bezel 6 Slide out the circuit board with the front panel intact 7 Remo...

Page 18: ...ard with the front panel intact 7 Remove the ba ery by sliding it forward toward the slot and replace with a 3 V bu on cell ba ery type BR1225 CR1225 or similar 8 Slide the circuit board back into the...

Page 19: ...e selectable AC cuto frequency 3 dB 0 5 Hz 0 1 dB 4 6 Hz max Input range 5 V AC voltage full scale range 5 Vpk min 5 1 Vpk typ 5 2 Vpk max Common mode voltage range AI to earth ground 2 V max IEPE exc...

Page 20: ...voltage range Vcommon mode Vbias Vfull scale must be 0 to 19 where Vcommon mode is the common mode voltage applied to the WebDAQ 504 Vbias is the bias voltage of the IEPE sensor Vfull scale is the ful...

Page 21: ...nput mode Pull up con gura on Each bit is pulled up to 5 V with a 100 k resistor Input frequency range DC 10 kHz Note 3 Input high voltage threshold 1 9 V min 3 6 V max Input low voltage threshold 2 3...

Page 22: ...nego ated Connector RJ 45 8 posi on Cable length 100 meters 328 feet max Addi onal parameters HP Auto MDIX support Network con gura on Table 7 Network con gura on Parameter Speci ca on Network IP con...

Page 23: ...derside of the device Note 5 When factory defaults are restored any shared folders or mapped drives are reset Processor Memory Table 9 Processor memory speci ca ons Parameter Speci ca on Microprocesso...

Page 24: ...MEDIA Blinking yellow A job is con gured to log to external media but the job hasn t started yet push bu on will not let you eject media Steady yellow External media is inserted into the SD card slot...

Page 25: ...on bu on FUNC Dual func on Ejects or safely unmounts all removable media for safe removal from the device default This func on is disabled when a job that uses external media is running The MEDIA LED...

Page 26: ...Parameter Speci ca on Dimensions L W H 158 8 146 1 38 1 mm 6 25 5 75 1 50 in 178 8 146 1 38 1 mm 7 04 5 75 1 50 in includes BNC connectors Weight 680 g 1 50 lb Shock and vibra on Table 18 Shock and v...

Page 27: ...by regulated low voltage sources and electronics Cau on Do not connect the WebDAQ 504 to signals or use for measurements within Measurement Categories II III or IV Signal connector Table 21 Screw term...

Page 28: ...w Voltage Direc ve 2014 35 EU RoHS Direc ve 2011 65 EU Conformity is assessed in accordance to the following standards EMC Emissions EN 61326 1 2013 IEC 61326 1 2012 Class A EN 55011 2009 A1 2010 IEC...

Page 29: ...Measurement Compu ng Corpora on 10 Commerce Way Norton MA 02766 Phone 508 946 5100 mccdaq com support WebDAQ 504 User s Guide page 29...

Reviews: