Chapter 8
128
4.
In the Reference Source box, select the reference that you are using (
Internal
or
External
;
Internal is the default).
5.
In the Auto Calibration box, click
Go
.
The bipolar (zero and full-scale) and unipolar (zero and full-scale) ranges are automatically
calibrated, and the calibration values are displayed. The bipolar readings should be within 0.001 V;
the unipolar readings should be within 0.0005 V.
6.
Click
OK
.
7.
In the Range box, select
PGH
Zero
.
8.
If the displayed value is not 0.0000 V (within 0.001 V), continue with
; otherwise, click
Quit
when you are finished calibrating
the analog input circuitry.
Once you have finished this procedure, continue with
“Calibrating the Analog Output
Note:
If you are not satisfied with the analog input calibration, you can load the factory
default settings stored in the EEPROM by clicking
Restore
in the Factory Settings box.
Using the Manual Calibration Procedure
If you want to manually calibrate the analog input circuitry instead of auto-calibrating it, do
the following:
1.
From the main menu of the DT3034 Calibration Utility, click
Configure
, and then
Board
.
2.
Select the name of the DT3034 board to configure from the combo box, and then click
OK
.
3.
From the main menu of the DT3034 Calibration Utility, click
Calibrate.
4.
Click
A/D
.
5.
In the Reference Source box, select the reference that you are using (
Internal
or
External
;
Internal is the default).
6.
In the Range box, select
Bipolar
, and then
Zero
.
7.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads 0.0000 V (within 0.001 V).
8.
In the Range box, select
Bipolar
, and then
+FS
(for full-scale).
9.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads 5 V with the internal reference or +9.3750 V with the external reference (within
0.001 V).
10.
In the Range box, select
Unipolar
, and then
+FS
(for full-scale).
11.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads 5 V with the internal reference or +9.3750 V with the external reference (within
0.0005 V).
12.
In the Range box, select
Unipolar
, and then
Zero
.
Summary of Contents for Data Translation DT3034
Page 2: ...DT3034 User s UM 22359 M Manual Title Page...
Page 5: ......
Page 11: ...Contents 10...
Page 15: ...About this Manual 14...
Page 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 20...
Page 22: ...Part 1 Getting Started...
Page 23: ......
Page 33: ...Chapter 2 32...
Page 39: ...Chapter 3 38...
Page 68: ...Part 2 Using Your Board...
Page 69: ......
Page 111: ...Chapter 6 110...
Page 123: ...Chapter 7 122...
Page 135: ...Chapter 8 134...
Page 141: ...Chapter 9 140...
Page 153: ...Appendix A 152...
Page 154: ...153 B Connector Pin Assignments...