Verifying the Operation of a DT3010 Series Board
95
Testing Single-Value Analog Output
To verify that the board can output a single analog output value, do the following:
1.
Connect an oscilloscope or voltmeter to DAC0 on the board. Refer to
for an
example of how to connect analog output signals.
2.
In the Quick DataAcq application, choose
Single Analog Output
from the
Control
menu.
3.
Select the appropriate DT3010 Series board from the Board list box.
4.
In the Channel list box, select analog output channel 0.
5.
In the Range list box, select the output range of DAC0.
The default is ±10 V.
6.
Enter an output value, or use the slider to select a value to output from DAC0.
7.
Click
Send
to output a single value from DAC0.
The application displays the output value on the screen in both text and graphical form.