Chapter 8
170
To calibrate DAC1 using the internal ADC, do the following:
1.
Connect Analog Out 1+ to Analog In 3.
2.
Connect Analog Out Return to Analog In 3 Return.
For the DT3010, DT3010-268, DT3010/32, DT3010/32-268 board, follow the instructions on
. For the DT3016 board, follow the instructions on
.
Configuring for an External Meter
To calibrate DAC0 using an external voltage meter, do the following:
1.
Connect Analog Out 0+ to the positive side of the precision voltage meter.
2.
Connect Analog Out 0 Return to the negative side of the precision voltage meter.
Signal
DT740 Screw
Terminal
STP268 Screw
Terminal on J1
STP268-EC
Screw Terminal
Analog Output 1+
TB43
TB43
TB16
Analog Input 3
TB7
TB7
TB31
Signal
DT740 Screw
Terminal
STP268 Screw
Terminal on J1
STP268-EC
Screw Terminal
Analog Output 1+ Return
TB44
TB42
TB51
Analog Input 3 Return
TB8
TB8
TB65
Signal
DT740 Screw
Terminal
STP268 Screw
Terminal on J1
STP268-EC
Screw Terminal
Analog Output 0+
TB41
TB41
TB17
Signal
DT740 Screw
Terminal
STP268 Screw
Terminal on J1
STP268-EC
Screw Terminal
Analog Output 0 Return
TB42
TB42
TB51
Summary of Contents for Data Translation DT3010 Series
Page 2: ...DT3010 Series UM 16866 V User s Manual Title Page ...
Page 5: ......
Page 15: ...About this Manual 14 ...
Page 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 21 ...
Page 23: ...Chapter 1 22 ...
Page 24: ...Part 1 Getting Started ...
Page 25: ......
Page 37: ...Chapter 2 36 ...
Page 49: ...Chapter 3 48 ...
Page 91: ...Chapter 4 90 ...
Page 102: ...Part 2 Using Your Board ...
Page 103: ......
Page 147: ...Chapter 6 146 ...
Page 159: ...Chapter 7 158 ...
Page 181: ...Chapter 9 180 ...
Page 193: ...Appendix A 192 ...
Page 213: ...Appendix B 212 ...
Page 231: ...Index 230 ...