Cleaning and service warnings and cautions
Do not mix disinfecting solutions (e.g., bleach and ammonia), as toxic gases may
result.
Make sure the ANI MOC-9 is installed outside the liquid projections hazard zone, e.g.
Perfusion bag.
Do not autoclave the ANI MOC-9. Autoclaving will seriously damage both components.
Compliance warnings and cautions
When using electro-convulsive therapy (ECT) equipment during ANI monitoring: place
ECT sensors as far away as possible from the ANI sensor to minimize the effect of
interference. Some ECT equipment may interfere with the ANI MOC-9 signal. Check
for equipment compatibility during patient setup.
Using accessories and cables other than specified or provided by the manufacturer of
the ANI MOC-9 (Mdoloris Medical Systems) may result in increased electromagnetic
emissions or decreased electromagnetic immunity of the ANI MOC-9 and may result in
an inappropriate operation.
The ANI MOC-9 complies with the electromagnetic compatibility requirements of EN
60601-1-2. Operation of this device may affect or be affected by other equipment in
the vicinity due to electromagnetic interference (EMI). If this occurs:
• increase separation between devices
• change the orientation of device cabling
• plug devices into separate outlets
• contact your Mdoloris Medical Systems representative.
The characteristics of ANI MOC-9 emissions allow it to be used in industrial areas and
hospitals (CISPR 11 class A). When used in residential environments (for which
CISPR 11 class B is required), ANI MOC-9 cannot guarantee provision of adequate