MD/PRD/IN16.ANIMOC9 V.02 - 12/03/2020
25 | 30
Chapter
7: Service and Maintenance
Cleaning and disinfection
Cleaning of the ANI MOC-9 should be performed at regular intervals or in accordance with
hospital, as well as local and governmental regulations.
See
Warning and caution instructions
section.
ANI MOC-9 is a reusable instrument which is supplied non-sterile.
Clean any blood or liquid spill on module immediately. Dried blood is very difficult to remove.
Use the maker’s wipes called “linget’ Anios” wipes (impregnated with ethanol) or “Wip’Anios”
wipes (impregnated with
Didecyldimethylammonium chloride) or
ethanol-based wipes.
After cleaning, allow it to dry completely. Residual moisture inside the connector may affect
the monitoring performance.
General maintenance
Safety tests should be done by qualified personnel only. Safety checks should be performed
at regular intervals or in, accordance with hospital, as well as local and government
regulations.
The following is a checklist for the general maintenance of the ANI MOC-9:
•
Visually inspect equipment for functional or structural damage, including poor seals,
cracks, damaged springs, etc.
•
Visually inspect cable, connector, and connector pins for signs of damage or wear,
•
Visually inspect product identification labels to ensure they are clear and legible,
•
System check and leakage current check according to the 62353 standard.
However leakage current must be checked systematically after every blood or liquid spill, or
immediately after a major surge in the electrical system.
Service Instructions
ANI MOC-9 has no customer serviceable parts.
Attempting to service ANI MOC-9 Module will
avoid the warranty.
Repair Policy
Mdoloris Medical Systems or an authorized Service Department must perform warranty
repair and service. Do not use malfunctioning equipment. Have the instrument repaired.