MD/PRD/IN16.ANIMOC9 V.02 - 12/03/2020
21 | 30
Chapter
5: Trouble shooting
To troubleshoot issues with Root, see the Operator's Manual for Root. To troubleshoot
issues with sensors, see the Instruction for Use of the ANI Sensor V1 PLUS.
Problem
Possible Cause
Solution
Bad computation quality
Cable is disconnected
Check the patient cable is
correctly connected to the
sensor.
Cable and connectors are
damaged
Check that the cables and
connectors are in good working
order.
Sensors are misplaced
Check that sensors are properly
placed along an imaginary line
through the heart (acquisition of
an electrical QRS axis).
Too much noise on the
signal
Check that any other device can
interfere.
ANI MOC-9 module is not
recognized
Power is not established
Module is not recognized
Unplug the module then restart
Root Monitor.