Cautions
: Do not autoclave the acquisition device or monitor. Autoclaving will seriously
damage both components.
Avoid liquid ingress with the connection cables. Contact of fluids with the connections cables
can interfere with the acquisition performance.
12.2
Leakage Current
The NIPE Monitor V1 is designed so that no periodic maintenance is required. However,
leakage current should be systematically checked, after spillage of blood or solutions, or after
a major surge in the house electrical system or at least once a year.
Preventive maintenance is recommended once a year: cable check, integrity of labels, system
check and leakage courant check according to the 62353 standard (for more details, refer to
the technical manual available from Mdoloris). However leakage current must be checked
systematically after every blood or liquid spill, or immediately after a major surge in the
electrical system.
Only Mdoloris Medical Systems qualified technicians and engineers or trained staff are
authorized to perform repairs and/or maintenance operations.
Only staff trained by Mdoloris Medical Systems can safely perform maintenance operations
(such as software updating or system recovery). However, the following elements may be
replaced or substituted by personnel untrained in technical maintenance (following the
manufacturer's instructions):
-
End User Cable;
-
Power cord;
-
Power unit.
13
Specifications, warranty and software license agreement
13.1
Specifications
Power requirements: 100-240 Volts
Consummation: 40 W
Current: < 3,4 A
Frequency: 50 Hz