Cleaning the monitor, power supply unitand end user cable:
clean any blood or liquid
spill on either the monitor or acquisition device immediately. Dried blood is very difficult to
remove. Use the maker’s wipes called “linget’ Anios” wipes or ethanol-based wipes.
After cleaning, allow it to dry completely. Residual moisture inside the connector may affect
the monitoring performance.
Do not use abrasive detergents to avoid scratching the screen.
WARNING
:
DO NOT MIX DISINFECTING SOLUTIONS AS TOXIC GASES MAY RESULT.
Caution:
Avoid liquid ingress with the connection cables. This could interfere with acquisition
performance.
11.2 Maintenance
The ANI Monitor V2 is designed so that no periodic maintenance is required.
Leakage current is a primary indicator of electrical shock hazard for personnel who touches
any of the device's outer surfaces. Static electricity tests have been conducted on our
monitor to ensure that the leakage current meets safety standards IEC 60601-1.
Preventive maintenance is recommended once a year: cable check, integrity of labels,
system check and leakage courant check according to the 62353 standard (for more details,
refer to the technical manual available from Mdoloris). However leakage current must be
checked systematically after every blood or liquid spill, or immediately after a major surge in
the electrical system.
Only Mdoloris Medical Systems qualified technicians and engineers or trained staff are
authorized to perform repairs and/or maintenance operations.
Only staff trained by Mdoloris Medical Systems can safely perform maintenance operations
(such as software updating or system recovery). However, the following elements may be
replaced or substituted by personnel untrained in technical maintenance (following the
manufacturer's instructions):
-
End User Cable;
-
Power cord;
-
AC-DC Power supply;
-
Pole clamp.