ANI Monitor V2 - Continuous analgesia monitoring system
MD/PRD/IN16.ANIV2 v.07 - 12 MAR 2020
P a g e
32 | 45
Real time data from the monitor can be retrieved
by connecting a computer to the “real-time
export” port on the right side of the monitor. The character string can be displayed by a
terminal type software application (to get the protocol available in the technical manual,
please contact your Mdoloris agent).
NOTE:
if a cable is inadvertently disconnected during a recording, reconnect the various
elements in the acquisition system as quickly as possible. When recording resumes it will be
considered by the program as a second complete recording. It is then crucial to indicate this
interruption in recording as an event so that this error is taken into account later when data is
processed.
7.8
Updating events
As explained above (see section 6.3), events corresponding to different clinical moments are
predefined in the system. The user can make these different events appear on the ANI trend.
Select
« Quit patient »
at the top of the screen during monitoring to stop the session and
return to the main menu. A message asks you to confirm.
Note: if you respond «Yes» you will not be able to return to the current patient. You will then
see the main menu.
To access and reconfigure the list of events, select
« Update events »
. The input screen for
events then appears: