McCrometer Dura Mag Quick Start Installation Manual Download Page 6

D

URA

 M

AG

4. 

Grounding

Sensor 

ground lug

12 gauge 

ground wire

12 gauge ground wire attached to mating 

flanges

Sensor ground lug

12 gauge 

ground wire

12 gauge ground wire attached to mating flanges

Sensor ground lug

12 gauge 

ground wire

Sensor 

ground 

lug

12 gauge 

ground 

wire

Grounding ring lug and 12 

gauge ground wire

Sensor 

ground lug

12 gauge 

ground wire

Optional earth

ground for

installations with 

building ground noise

Sensor 

ground lug

Grounding ring lug and 12 

gauge ground wire

Next, connect the two 

mating grounding rings 

with a 12 gauge wire or 

equivalent, attached to the 

grounding ring lugs, and 

connect them both to the 

earth ground. The Dura 

Mag should be electrically 

isolated from the pipeline.

2.  Sensor grounding for meters in an electrically noisy environment

If there is electrical noise in the fluid column or electrical current in the pipe, it can be minimized or eliminated using 

grounding rings or by grounding the pipeline. This applies to meters mated to conductive uncoated pipe.
Attach the provided 12 gauge wire, or equivalent, to the ground lug. Then connect the sensor ground lug to an earth 

ground point. 
Next, connect the two mating pipe sections with a 12 gauge wire or equivalent and connect them both to the earth 

ground. The Dura Mag should be electrically isolated from the pipeline.

3.  Sensor grounding for meters with 

minimal ground noise

Attach the provided 12 gauge wire, or 

equivalent, to the sensor ground lug. 

Then connect this sensor ground lug to 

an earth ground point.

1.  Preferred method of grounding

Dura Mag meters come standard with a set of grounding rings for use with the preferred method of grounding Dura Mag 

meters.
This method can be used for all installations, but it is required for non-conductive or internally coated pipe. When pipes are 

non-conductive, such as PVC or internally coated pipe, you must substitute direct grounding with grounding rings.
Attach the provided 12 gauge wire or equivalent to the sensor ground lug. Then connect this sensor ground lug to  an earth 

ground point; refer to NEC or local grounding regulations for wiring requirements in making this connection.

30122-61  Rev. 1.4 | 04OCT2021

Page 6

Summary of Contents for Dura Mag

Page 1: ...r removal of meters can result in serious injury or death Read the instructions in this guide on the proper procedures carefully Any person installing inspecting or maintaining a McCrometer flowmeter should have a working understanding of piping configurations and systems under pressure Before adjusting or removing any meter be certain the system has depressurized completely Be careful when liftin...

Page 2: ...e to the meter s junction box and the converter s external connection Flanged Meter Installation Install the Dura Mag flow meter inline between two flanged end pipes The flow meter may require grounding depending on the environment they are being installed in Refer to section 4 for a full description of grounding methods that are available Shown are customer s pipe ends and flanges See section 3 f...

Page 3: ...an electronics shed or environmental enclosure The sun shield should be oriented in a direction to reduce sun damage and ensure readability Mount the converter to a solid surface using four bolts This electronic unit is rated IP67 for temporary flooding 2 Remote Mount Converter Installation Installing Cables through Cable Glands and Conduit All electrical cables enter the converter through compres...

Page 4: ...ep the attached cap firmly screwed into the connector to insure a water tight seal Also keep the contacts in the cable connector clean and dry during assembly Standard cable gland Quick connect connectors Port assignment 1 Outputs 2 Outputs 3 Power 4 Coils 5 Electrodes 1 2 3 1 2 3 4 5 Cable Gland Assignment for Wiring Harnesses Meter mount view Remote mount view Terminal Block Diagram and Groundin...

Page 5: ... colors in the McCrometer SmartOutput interface cable are identified together in the top row of the table above Corresponding wire colors for transceivers from each compatible AMI vendor are identified in the columns under the top row Terminal Port Wire Color 8 4 Green Yellow 11 4 Red 14 4 Yellow Chassis Lug 4 Black Sensus Itron Badger Neptune Elster 31 Black Green Black Green Black 32 Red Black R...

Page 6: ...conductive uncoated pipe Attach the provided 12 gauge wire or equivalent to the ground lug Then connect the sensor ground lug to an earth ground point Next connect the two mating pipe sections with a 12 gauge wire or equivalent and connect them both to the earth ground The Dura Mag should be electrically isolated from the pipeline 3 Sensor grounding for meters with minimal ground noise Attach the ...

Page 7: ...nverterdisplayislightactivatedandrequiresaminimum amount of light to appear Environments where light is low such as in dimly lit buildings or outdoors after sundown may prevent the display from appearing when the lid is raised There is an optical sensor embedded in the display located under the McCrometer swirl logo at the lower left If the display does not appear a flashlight will provide suffici...

Page 8: ...t the internal wiring is correct in the converter for the coil wires ADC CLIP The meter signal exceeds the limit of the converter likely due to noise Check the meter grounding meets the IOM guidelines and identify any sources of noise HIGH HUM Humidity inside converter housing has exceeded limits Check for any loose connections on converter housing that could be creating a leak COIN LOW Datalogger...

Reviews: