A0 version

 

 

 

 

 

 

 

6.4

 

Run a treatment 

6.4.1

 

Start the treatment 

 
Use the BioFeedBack exercise module. 

 
Reminder: Taring Biofeedback 

The taring works automatically. However, it is possible to adjust the parameters calculated by 

the machine with the buttons

 and

 . 

 
Procedure for taring 

 

Place the probe or electrodes  

 

Start taring 

 

Ask the patient to make a sustained contraction (the gauges change automatically), 
then release the effort for a few seconds. 

 

The BioStim automatically sets the operating range 

 

Go to the next page by pressing the arrow (or wait for the end of the taring time) 

 

During the exercise, it will always be possible to manually adjust the level of the tare 
with the + and - buttons at the top right of the page 

 

 

 
 
 
 
 
 
 
 
 

It is possible for each channel to activate or not the negative biofeedback, either in the 
configuration page or during the taring. 
If the option is enabled, you can see the work below the base level. This is particularly useful if 
you are working on relaxation. 

  

 
 
 
 
 
 

In the + version, it is possible to adjust the level of negative BFB to be displayed during the 
exercise, using the slider in the right-hand panel: 
Zero Patient: the bottom of the screen corresponds to the minimum reached by the patient 
during the taring 
This setting allows you to erase the basic tone. 

Negative biofeedback 
zone 

Max Level 

Basic Toning 
Zero Patients 

Zero Biostim 

Max Level 

Zero Patients 

Zero Biostim 

Summary of Contents for BIOMOOV

Page 1: ...Page 1 on 20 DATE 13 04 2022 A0 version BIOMOOV Manual Proprioception sensor...

Page 2: ...been designed and manufactured for use in a therapeutic setting Use is reserved for professionals who have undergone the appropriate training If you have any problems or do not understand this manual...

Page 3: ...he sensor has no therapeutic claims Equipped with a software the BioMoov allows through a Bluetooth connection to the Biostim to transcribe the patient s activities on a proprioception board The BioMo...

Page 4: ...ce 11 6 1 1 Powering up starting up 11 6 1 2 Using Bluetooth 11 6 1 3 Warning screen 11 6 2 Choosing a treatment 12 6 3 Setting the parameters 12 6 4 Run a treatment 13 6 4 1 Start the treatment 13 6...

Page 5: ...ion are technically up to date at the time of marketing However we reserve the right to make changes to the device and its documentation without any obligation to update these documents In the case of...

Page 6: ...g instructions must be read for safe use of the device Recycling This device should be disposed of at an appropriate collection and recycling facility Consult the manufacturer Manufacturer Date of man...

Page 7: ...Operating temperature 0 C to 40 C Storage temperature 40 C to 70 C Operating relative humidity 30 to 75 Operating altitude 2000 meters 2 2 2 Technical characteristics of BioMoov Housing dimensions 5 8...

Page 8: ...rsonal use only The BioMoov is not in contact with the patient Indeed it is intended to be fixed only on a proprioception tray The use of accessories not recommended by the manufacturer does not engag...

Page 9: ...the unit CAUTION No modifications to the device are permitted It is strictly forbidden to open the housing of the device CAUTION This device complies with applicable electromagnetic compatibility sta...

Page 10: ...Check the contents against the packing list included with the documentation Check that the contents of the box are not damaged if you have any doubts about the integrity of the device or its accessor...

Page 11: ...wering up starting up Turn on the device with the on off button located on the device The BioMoov turns on and a LED flashes 6 1 2 Use of Bluetooth See setting specification 6 1 3 Warning screen When...

Page 12: ...menu If the Biostim is not connected to the PC the duration of the programs is limited to 1 minute When accessing the treatment and before launching it it is necessary to set up the tray Clicking on...

Page 13: ...g the exercise it will always be possible to manually adjust the level of the tare with the and buttons at the top right of the page It is possible for each channel to activate or not the negative bio...

Page 14: ...ring processing by selection on the screen or by pressing the buttons on the remote control optional The treatment can be stopped or restarted by selection on the touch screen or by pressing the keys...

Page 15: ...claims there are no side effects 8 Maintenance upkeep The BioMoov device is designed to last 5 years To guarantee the preservation of the device s performance throughout its life it is necessary to h...

Page 16: ...et ZA Route de Tence 43520 Le Mazet St Voy Tel 33 4 71 65 02 16 Fax 33 4 71 65 06 55 Email sav electroniquedumazet com Possible operating anomalies Description of the anomaly Possible causes Actions N...

Page 17: ...du Mazet and its authorized distributors of all responsibility for defects malfunctions damage injuries and the like The warranty is void if the operating instructions in this manual are not strictly...

Page 18: ...n 18 05 2022 14 Manufacturer lectronique du Mazet is a company located in the heart of the Massif Central Originally a simple manufacturer of electronic cards over the years it has developed its own b...

Page 19: ...Page 19 on 20 DATE 13 04 2022 A0 version Your dealer distributor ELECTRONIQUEDU MAZET ZA ROUTE DE TENCE 43520 LE MAZET SAINT VOY T l 33 4 71 65 02 16 Mail sav electroniquedumazet com...

Page 20: ...or its use maintenance care etc I have read the instruction manual and I have noted the warranty and service conditions In the event that Electronique du Mazet or its distributors do not receive this...

Reviews: