4–5
OPERATION
www.maxcessintl.com
MAGPOWR DTR-65
MI 850A349 1 D
Tension calibration
The DTR-65 allows either precision calibration or weightless
calibration. The Precision Calibration method is described
below. Weightless Calibration begins on page 4-8.
Calibration can only be performed when the password has not
locked the DTR-65. To unlock the DTR-65, see the Password
parameter P.00 on page 5-4.
Precision calibration
Use this method when applying a known weight with a rope in
the web path to the load cells for calibration. Typically this
should be 10% of the load cell rating or greater.
Enter the programming mode by pressing and holding [+] and
[-] for two seconds.
Use the [+] and/or [-] buttons to go to parameter P.01. Press
[
] to see the present load cell excitation voltage. Select either
5V or 7.5V. When using MAGPOWR load cells select 7.5V. Press
and hold [
] for two seconds (until the display starts flashing)
to store the changes.
Use the [+] and/or [-] buttons to go to "CAL1". Press [
] to
enter the CAL1 sub-menu.
Display now shows C1.1. Unload the load cell roller and press
[
]. The DTR-65 will now zero out the idler roll weight and
show a wait display. After zeroing, the DTR-65 will
automatically increment to the next parameter.
continued